![HBM CX27B Скачать руководство пользователя страница 130](http://html.mh-extra.com/html/hbm/cx27b/cx27b_operating-manual_2111759130.webp)
Modules and transducers
130
I3031-14.0
HBM: public
Quantum
X
Notice
The transducer excitation voltage can be set in the range
of 5 ... 24 V (as described in 6.1.2) is only available on
channels 1 ... 8.
On channels 9 ... 16 the supply voltage (10 ... 30 V) from
the module (e.g. 24 V) is output less approx. 1 V.
A current of max. 30 mA can be consumed for these
channels. The current limitation switches the transducer
excitation off if current consumption is higher.
8.10.2 MX1601B status display
The front panel of the universal amplifier has a system
LED and 16 connection LEDs. The system LED indicates
the status of the device, the connection LEDs the states
of the individual connections.
System
LED
Connection LED
Connection LED
Fig. 8.17
MX1601B front view