Transducer connection
Quantum
X
I3031-14.0
HBM: public
145
9.8
Connecting transducers with double
shield technique
5
8
10
7
Measurement signal (+)
Excitation (+)
Sense lead (-)
Sense lead (+)
Excitation (-)
Measurement signal (-)
Cable shield
2
3
Hsg.
4
9
1
2
4
3
RB / 2 (to transducer)
Hsg. = Housing
1
5
6
10
11
15
For amplifiers MX840B, MX440A and MX410B, we re
commend this connection technique for very small meas
urement ranges, particularly in environments subject to
interference and when using long cables.
This applies for all bridge connections.
With cable lengths >50 m, a resistor with half the value of
the bridge resistance (RB/2) must be connected in each
sense lead of the transducer.