![HBK CB1010 Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/hbk/cb1010/cb1010_operating-manual_2111690038.webp)
CANHEAD
CONNECTING TO MGCPLUS
36
8
CONNECTING TO MGCPLUS
MGCplus
CANHEAD
(from the back)
Termination resistor
or more CANHEAD
1)
The AP74 connection board in the MGCplus comes with an integrated termination resistor
The documentation for the CANHEAD system with ML74B/AP74 is located on the
MGCplus system CD.
8.1
How many modules are possible at what cable length?
The strain gages or sensors connected to the CANHEAD modules are fed with the supply
voltage through the bus line. Therefore, the bus length is limited, if a major number of
CANHEADs is connected to one bus line.
In the most unfavorable case, all CANHEAD modules are arranged in close proximity to
each other. Using Devicenet Thin Media cable and taking into account the ambient tem
perature results in maximum bus lengths as specified in the table (see table specifying
the maximum bus lengths in the technical Datasheet).
Содержание CB1010
Страница 1: ...CANHEAD ENGLISH DEUTSCH Operating Manual Bedienungsanleitung...
Страница 3: ...CANHEAD ENGLISH DEUTSCH Operating Manual...
Страница 40: ...CANHEAD CONFIGURING WITH MGCPLUS 38 S Serial number and bridge type of the connected module...
Страница 52: ...CANHEAD CANHEAD ENHANCEMENTS 50...
Страница 53: ...CANHEAD ENGLISH DEUTSCH Bedienungsanleitung...
Страница 91: ...39 CANHEAD KONFIGURIEREN MIT MGCPLUS S Seriennummer und Br ckentyp des angeschlossenen Moduls...
Страница 103: ...51 CANHEAD CANHEAD ERWEITERUNGEN...
Страница 104: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A01185 13 X00 00 7 0102 0010...