![HBK CB1010 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/hbk/cb1010/cb1010_operating-manual_2111690037.webp)
35
CANHEAD
CONNECTION
ing measuring point; should, for example, an SG become disconnected from mea
suring point 8, the red LED will flash 8 times.
Hexa‐
deci
mal
Error number
(decimal
value)
Error
Remedy
1
1
internal bus error
Repair situation
2
2
Error during hardware testing
Repair situation
4
4
Setup data (system variables) cannot
be loaded; e.g. bridge excitation voltage
Enter the correct
variable
8
8
Setup data (system constants) cannot
be loaded; e.g. calibration values
Enter the correct
constant
10
16
Errors in certain parameters (e.g. CAN
address) in the system variant or sys
tem constant
Enter the correct
parameter
20
32
Wrong command sent from ML74B to
CANHEAD (STO)
Repair situation
40
64
Execution error
Repair situation
80
128
Measuring point error
Check the measur
ing point
100
256
Synchronization error: gate array error
Repair situation
200
512
Synchronization error: bridge excitation
voltage
Repair situation
1)
when the error keeps occurring
Examples
LED flashes
7
times: errors 1+2+4:
S
Internal bus error, hardware test failure and wrong variables entered
LED flashes
20
times: errors 16+4:
S
Wrong parameters (setup data) entered
Содержание CB1010
Страница 1: ...CANHEAD ENGLISH DEUTSCH Operating Manual Bedienungsanleitung...
Страница 3: ...CANHEAD ENGLISH DEUTSCH Operating Manual...
Страница 40: ...CANHEAD CONFIGURING WITH MGCPLUS 38 S Serial number and bridge type of the connected module...
Страница 52: ...CANHEAD CANHEAD ENHANCEMENTS 50...
Страница 53: ...CANHEAD ENGLISH DEUTSCH Bedienungsanleitung...
Страница 91: ...39 CANHEAD KONFIGURIEREN MIT MGCPLUS S Seriennummer und Br ckentyp des angeschlossenen Moduls...
Страница 103: ...51 CANHEAD CANHEAD ERWEITERUNGEN...
Страница 104: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A01185 13 X00 00 7 0102 0010...