Manual  Conpreva

Default setting of the timer

relax:

15s

tense:

 8s

current:

 5s

break:

 10s

The times can be changed by the physician or therapist between 1s and
200s.  The change of the training phases is indicated by an acoustic signal.

EMC

Emission: EN 55011 

class A

EN 55014 for click noise

Immunity: ESD 

(IEC 61000-4-2)

2+4+6kV Contact, 2+4+8kV Air

Radiated Field 

(IEC 61000-4-3)

3V/m , 80-2500 MHz, AM: 80% , 1kHz or 2 Hz, Modulation Frequency;
10V/m from 80-2500MHz

BURST 

(IEC 61000-4-4) 2 kV

SURGE 

(IEC 61000-4-5)

0.5kV, 1kV differential mode
0.5kV, 1kV, 2kV common mode

Conducted RF Immunity 

(IEC 61000-4-6)

3V, 80% AM, 0.15-80 MHz, 1kHz or 2Hz Modulation Frequency;
different startfreq for battery powered systems

Magnetic Fields 

(IEC 61000-4-8) 50 Hz, 10A/m

AC variations 

(IEC 61000-4-11)

Voltage Test Level 0 %, 40 and 70 %;
Duration (Periods) 0.5, 5, 25; Voltage Interruption 5 sec

Security options

• 

Prevention of high regulating of the current outside  the stimulation phase

• 

soft increase/ decrease of the current

• 

Electrode alarm automatic limitation and adjustment of the current during

contact reduction

• 

Protection class: II; IP 20

Output

Current:  adjustable: mono/ biphasic
max. current:

   250 mA on 500 Ohm on 27Hz

effect. current:

   9,9 mA on 1000 Ohm (on 9,9V)
   13,2 mA on 500 Ohm (on 6,8V)

pulsewidth:

   adjustable: 40 

μ

s up to 500

μ

s

frequency:

   adjustable: 2 Hz up to 200 Hz

max. amplitude on 500 Ohm: ±160 V on 5 Hz
max. amplitude on open output: ±170 V
max. energy of a single pulse on a load impedance of 500 Ohm: maximal
2 mJ
modulation: amplitude- and frequency modulation

Pulse pattern

Technical data

Technical data

40 to

 500µs

1/f-640µs to

1/f-1100µs

100 to
560µs

1/f

U

[V]

t

150

-150

0

40 to

 500µs

Содержание Conpreva

Страница 1: ...ck training 13 Electrical stimulation training 15 The combined training Biofeedback electrical stimulation 16 EMG triggered electrical stimulation 18 operation mode pressure 20 preparation 20 set the...

Страница 2: ...triggered electirical stimulation with adjustable pulse width is available The therapeutic use of the Conpreva is based upon a learning strategy whose origin lies within the psychology which aim for r...

Страница 3: ...ase read through this manual completly before you start using the device The device may only be used with the accessories provided by the producer Every use that excesses of this is not allowed Contra...

Страница 4: ...s that emit strong electromagnetic fields to prevent measuring errors In case of allergies towards polymer foam gold and silver please inform your doctor before you start the training Please inform yo...

Страница 5: ...g outside electromagnetic fields could cause measuring errors Clean thoroughly the skin in the places on which you place the electrodes The skin must be perfectly dry and released from hair Use contac...

Страница 6: ...urrent during the exercise Exercise with electrical stimulation uses only channel 1 for measurement so only one emg scale is shown Course of exercise Course of exercise Flowing current in mA and graph...

Страница 7: ...G is overriden correct during maximum tense within the upper range wrong EMG is adjusted too roughly remaining exercise duration current phase battery state EMG range remaining duration of the current...

Страница 8: ...hold it loose and relaxed Electrical stimulation training Accomplish the steps described in preparation Program selection and EMG setting Follow the commands of the voice to relax the musculature The...

Страница 9: ...nimation does not please you you have the possibility to select between different animations Press in addition during the tense or relax phase the Resume button Five further diagrams stand to the sele...

Страница 10: ...increase the current at any time In order to increase the current you must turn first the turn switch thus downward against the clockwise direction If the lock is open you can increase the current The...

Страница 11: ...the EMG tension on the first channel This mode of operation is suitable to measure the pressure in the rectum and the muscle tension of the abdominal muscles at the same time Also the combination of a...

Страница 12: ...l the plug of the sensor cable from the sensor socket and remove the sticking electrodes Keep these for reuse and take care of them as described under maintenance If used remove the headphones Chargin...

Страница 13: ...afety relevant controls in accordance with MedBetreibV 6 prescribed product according with annex 1 of the regulation Attention keep the plugs dry Maintenance Cleaning the sensor Clean the sensor after...

Страница 14: ...tteries or power supply is defect contact the manu factorer on off button doesn t work rechargeable batter ies are fully dis charged charge the batteies Debugging error Reason solution plug of the ada...

Страница 15: ...ters which are important for the therapy e g all times or the frequencies with which the patient has to train Stan dard disease patterns are pre defined The optimal according to scientific knowledge t...

Страница 16: ...grams are available under the menu entry torso To adjust the exercise duration you have to choose Exercise duration within System settings therapy spec parameters The choosen exercise duration does ap...

Страница 17: ...he following way automatically during the training phases the optimal EMG range is constantly selected semiautomatically during the actual EMG setting the patient strongly tense up 5 seconds The optim...

Страница 18: ...ume key Enter the surname of the patient Confirm the input by pressing the resume button Now you can check the entered surname Press the resume button Enter successively first name address date of bir...

Страница 19: ...divided in minutes i e 01 12 means one minute and 12 seconds The scaling of the abscissa orients itself at the maximum EMG value of the respective exercise in V The tense phase is brown represented th...

Страница 20: ...press the turn switch for more then one second Hold it down and turn the turn switch to set the new volume You can change the volume every time excepting the stimulation phase Example Music volume se...

Страница 21: ...rate and Trigger limit with appedant time limit If the trigger automatic is activated the threshold moves down if the patient doesn t achieve the given threshold value The update rate determines the...

Страница 22: ...t for headphones Multilink for connection with DMA or charging device LG PTA med Power supply NiMH recharable batteries size AA 2500 mAh 1 2 V charging device LG PTA 230V 50Hz Technical data Maintenan...

Страница 23: ...8 50 Hz 10A m AC variations IEC 61000 4 11 Voltage Test Level 0 40 and 70 Duration Periods 0 5 5 25 Voltage Interruption 5 sec Security options Prevention of high regulating of the current outside the...

Страница 24: ...es of each exercise are documented so that the execution of the training as well as the adherence to the training defaults can be examined For each patient an allocation of the training data is possib...

Страница 25: ...er authorized by the company Any modifications or repairs performed by unauthorized centers or groups will void this warranty Any malfunction or failure in the product caused by product misuse includi...

Отзывы: