20
21
Manual Conpreva
operation mode pressure
preparation
•
connect the sensor plug to the according socket at
the
DMA
•
after this connect the plug of the
DMA
to the
multilink of the Conpreva
•
insert the pressure sensor
•
In the case of combinated measurement (pressure
+ EMG) follow the instructions described in
Preparation
•
three diagnosis modes and one exerise is
available- select one with the
turn switch
and press
the
resume
key
set the pre pressure
Depending to the sensor and/or indication it is
necessary to adjust the pre pressure (basic pressure).
This pre pressure is the zero-line in the following
exercise. The Conpreva can measure a pressure of
250 mBar over the normal pressure (ambient pressure).
After you setted the pressure, press the
resume
key.
Depending to the selected program you come directly
to the
EMG setting
or direct into the exercise or
diagnosis.
start the diagnosis
diagnosis pressure
This program measures only the pressure. The zero-
line of the ordinate corresponds to the adjusted pre
pressure. The pressure can become negative, if the
musculature is more relaxed than in the pre pressure
setting. All diagnostic programs only record data, i.e.
the Conpreva does not give any instructions like tense
up or an easing.
diagnosis pr 1 EMG
This program measures the pressure as well as the
EMG tension on the first channel. This mode of operation
is suitable to measure the pressure in the rectum and
the muscle tension of the abdominal muscles at the
same time. Also the combination of a special sensor
(AS2000 P) with a balloon is possible. Therefor the
display is splitted. The pressure is displayed in a dia-
gram in the upper area. The EMG tension is displayed
in the colored scale. Additionally the EMG tension is
composed.
diagnosis pr 2 EMG
This program measures the pressure and the EMG
tension on both channels. This mode of operation is
suitable, to measure in combination with the AS2000
P the pressure in the rectum, the EMG tension of the
sphincter and the muscle tension of the abdominal
muscles at the same time. The pressure is shown in a
diagram within the upper area. The EMG tension is
shown on two colored scales. The left scale
corresponds to sensor socket 1, the right scale
according to sensor socket 2. Additionally the EMG
tension is composed. The sound always corresponds
thereby to the maximum of the two EMG scales.
training pressure
This exercise uses a fixed tense up time of 3s and a
relax time of 10s. The measuring range is scaled
automatically to the maximum. The right representation
shows a surface diagram. Hold the curve in the grey
area. The tense up phase is brown indicated, the relax
phase green. In addition a flower, a scale and a further
surface diagram are available while pressing the
resume key in while the exercise.
Course of exercise
Course of exercise