42
43
Manual Conpreva
Technical data
Modes of operation
•
EMG-biofeedback-training
• Electrical stimulation
• Combined training with biofeedback and electrical
stimulation
• EMG-triggered electrical stimulation
•
Manometric-biofeedback-training
•
Manometric-diagnosis
•
Manometric-diagnosis with EMG-biofeedback 1 or 2 channel
EMG-voltage measurement range
•
0 - 300
μ
V with a resolution of 0,01
μ
V .. 0,25
μ
V
•
displayed as absolute value and graphical
•
EMG-range can set manully, semi-automatic and fully automatic
pressure measurement range
• 0 - 250 mBar above normal pressure
Display
240*320 pixel multi color display
Sockets
Two three pole connection plugs on the device’s top for the connection of
the electrodes. 5,5 mm socket for the connection of the battery charger
and 4 pole socket for the connection to the PC over a USB cable connection
(5000V examined/optically decouples). 3.5 mm of socket for headphones.
Multilink for connection with DMA or charging device LG PTA med
Power supply
NiMH recharable batteries size AA 2500 mAh, 1,2 V
charging device LG PTA (230V, 50Hz)
Technical data
Maintenance
The menu
Maintenance
includes functionalities to set
the device into
primitive state
. Note that thereby all
patient data, exercise data and therapy settings are
lost. Try first:
•
Put first an adequate pin approx. 5 millimeters into
the
reset button
at the right device side. After a
few seconds the device is ready for use and all
settings are stored.
•
If the error still occurres, select the option
primiti-
ve state
If the error is still present after this, a defect is present.
Please send the device back to the manufacturer.
Activate
Delete all exercises
if the exercise memory
is full and there is no PC to read out the exercise data.
The options
Change application
and
Calibration
need
the administrator password, so they are non-
accessible.
To set the time rotate
the turn switch
. The input
sequence is: Hour, minute, year, month and finally
day.
Further informations
Reset button