Therapy settings

Inside the menu 

therapy settings

 extensive functions

are available, with whose assistance you can select
the suitable therapy for each patient. You have the
choice between these menu options:

• head / Extremities / torso (pressure)
• manually configuration
• general exercises

Over all up to seven different exercises are possible,
from which the patient will be able to choose one at
time. Therefore you can create a training schedule that
will be shown when the device starts. This training
schedule can contain up to seven different exercises,
consisting of fixed exercises, general exercises and
manually configurations. If you have already activated
seven exercises, you will be unable to activate any
further ones. First you must deactivate an active one.

Head / extremities / torso

Here you can find diagnoses that can be treaten with
the device. The numerous headwords, for example
migraine, describe the best therapy parameter in form
of fixed programs, designed after the current scientific
knowledge. After choosing one of these programsa
green check mark is visible behind the entry. Pressing
on this entry again will deactivate it - the green check
mark disappears.
Additional to this you will be asked to choose a position
for the electrodes. This specific position for the
electrodes will be displayed everytime the exercise is
being started.
The pressure and incontinence programs are available
under the menu-entry torso.

To adjust the exercise duration you have to choose

Exercise duration 

within

 System settings/therapy

spec. parameters

. The choosen exercise duration

does apply for all fixed exercises. You can set the 

EMG-

setting

 also in the menu 

System settings/Therapy

spec. parameters/EMG-setting

.

Manually configuration

You can create six own configurations with your own
identification. Then the exercise programs can be
individually arranged for each patient, by activating or
deactivating the exercises.

Choose the 

Operation mode

 with wich the patient

have to work

Adjust the 

frequency 1

 and 

frequency 2

. You must

adjust these two frequencies, since these
alternate after each current phase. This makes
possible to work with changing current forms in an
exercise - e.g. you can work in the one phase with
a frequency modulation, in the other phase with a
fixed frequency.  If you would like to use only one
frequency, adjust for both frequencies the same
value. In order to activate the frequency
modulation, select 

frequency modulation

 instead

of 

fixed frequency

.  The modulation is processed

within 3 seconds from 4 Hz to 70 Hz.
In addition you can specify whether the device skip
the 

current setting

 and a set a 

fixed current

, or the

patient can adjust the current. If you select 

fixed

current

 you jump into the current setting. Note that

you should specify the frequency and pulse width
before, since the effective current and thus feeling
of the current depend also on the frequency. The
pulse width is adjustable between 40

μ

s and 500

μ

s

Further informations

Further informations

Содержание Conpreva

Страница 1: ...ck training 13 Electrical stimulation training 15 The combined training Biofeedback electrical stimulation 16 EMG triggered electrical stimulation 18 operation mode pressure 20 preparation 20 set the...

Страница 2: ...triggered electirical stimulation with adjustable pulse width is available The therapeutic use of the Conpreva is based upon a learning strategy whose origin lies within the psychology which aim for r...

Страница 3: ...ase read through this manual completly before you start using the device The device may only be used with the accessories provided by the producer Every use that excesses of this is not allowed Contra...

Страница 4: ...s that emit strong electromagnetic fields to prevent measuring errors In case of allergies towards polymer foam gold and silver please inform your doctor before you start the training Please inform yo...

Страница 5: ...g outside electromagnetic fields could cause measuring errors Clean thoroughly the skin in the places on which you place the electrodes The skin must be perfectly dry and released from hair Use contac...

Страница 6: ...urrent during the exercise Exercise with electrical stimulation uses only channel 1 for measurement so only one emg scale is shown Course of exercise Course of exercise Flowing current in mA and graph...

Страница 7: ...G is overriden correct during maximum tense within the upper range wrong EMG is adjusted too roughly remaining exercise duration current phase battery state EMG range remaining duration of the current...

Страница 8: ...hold it loose and relaxed Electrical stimulation training Accomplish the steps described in preparation Program selection and EMG setting Follow the commands of the voice to relax the musculature The...

Страница 9: ...nimation does not please you you have the possibility to select between different animations Press in addition during the tense or relax phase the Resume button Five further diagrams stand to the sele...

Страница 10: ...increase the current at any time In order to increase the current you must turn first the turn switch thus downward against the clockwise direction If the lock is open you can increase the current The...

Страница 11: ...the EMG tension on the first channel This mode of operation is suitable to measure the pressure in the rectum and the muscle tension of the abdominal muscles at the same time Also the combination of a...

Страница 12: ...l the plug of the sensor cable from the sensor socket and remove the sticking electrodes Keep these for reuse and take care of them as described under maintenance If used remove the headphones Chargin...

Страница 13: ...afety relevant controls in accordance with MedBetreibV 6 prescribed product according with annex 1 of the regulation Attention keep the plugs dry Maintenance Cleaning the sensor Clean the sensor after...

Страница 14: ...tteries or power supply is defect contact the manu factorer on off button doesn t work rechargeable batter ies are fully dis charged charge the batteies Debugging error Reason solution plug of the ada...

Страница 15: ...ters which are important for the therapy e g all times or the frequencies with which the patient has to train Stan dard disease patterns are pre defined The optimal according to scientific knowledge t...

Страница 16: ...grams are available under the menu entry torso To adjust the exercise duration you have to choose Exercise duration within System settings therapy spec parameters The choosen exercise duration does ap...

Страница 17: ...he following way automatically during the training phases the optimal EMG range is constantly selected semiautomatically during the actual EMG setting the patient strongly tense up 5 seconds The optim...

Страница 18: ...ume key Enter the surname of the patient Confirm the input by pressing the resume button Now you can check the entered surname Press the resume button Enter successively first name address date of bir...

Страница 19: ...divided in minutes i e 01 12 means one minute and 12 seconds The scaling of the abscissa orients itself at the maximum EMG value of the respective exercise in V The tense phase is brown represented th...

Страница 20: ...press the turn switch for more then one second Hold it down and turn the turn switch to set the new volume You can change the volume every time excepting the stimulation phase Example Music volume se...

Страница 21: ...rate and Trigger limit with appedant time limit If the trigger automatic is activated the threshold moves down if the patient doesn t achieve the given threshold value The update rate determines the...

Страница 22: ...t for headphones Multilink for connection with DMA or charging device LG PTA med Power supply NiMH recharable batteries size AA 2500 mAh 1 2 V charging device LG PTA 230V 50Hz Technical data Maintenan...

Страница 23: ...8 50 Hz 10A m AC variations IEC 61000 4 11 Voltage Test Level 0 40 and 70 Duration Periods 0 5 5 25 Voltage Interruption 5 sec Security options Prevention of high regulating of the current outside the...

Страница 24: ...es of each exercise are documented so that the execution of the training as well as the adherence to the training defaults can be examined For each patient an allocation of the training data is possib...

Страница 25: ...er authorized by the company Any modifications or repairs performed by unauthorized centers or groups will void this warranty Any malfunction or failure in the product caused by product misuse includi...

Отзывы: