Haynl-Elektronik Conpreva Скачать руководство пользователя страница 14

26

27

Manual  Conpreva

Debugging

Debugging

error 

reason 

solution 

no connection be-
tween cable and  

sensor socket

 

check the plug and 
the 

sensor socket

  

Insufficient skin con-
tact because the 
electrode is partial 
or completely flaked 

use new electrodes 
or humidify the stick-
ing side 

no EMG-signal  

EMG-range is too 
large 

select smaller EMG 
range (see chapter 
„EMG-setting“ 

non EMG-signal – 
(see above) 

(see above) 

EMG-range is too 
small 

select larger EMG 
range (watch chap-
ter „EMG-setting“ 

overrided 
EMG-signal 
(scale off) 

strong electromag-
netic fields 

change into a „cle-
an“ room 

the device 
doesn’t charge  

wrong connection 
between 

charging 

plug

 and 

charging 

socket

 

check the plug and 
the 

charging socket

 

the fully char-
ged battery 
works shorter 
than 2 hours 

the rechargeable 
batteries or power 
supply is defect 

contact the manu-
factorer 

on/off button 
doesn’t work 

rechargeable batter-
ies are fully dis-
charged  

charge the batteies  

 

Debugging

error 

Reason 

solution 

plug of the adaptor 
cable is incorrectly 
connected with the 

sensor socket

 

check the plug in 
the 

sensor socket

 

insufficient skin con-
tact because the 
electrode is partial 
or completely flaked 

use new electro-
des or humudify 
the sticking side of 
the electrode 

the following 
message is 
shown 
 

 

device is on the po-
wer limit (high skin 
resistance) 

use contact gel, 
new electrodes or 
needle electrodes 

I can’t increase 
the current 

the equipment has a 
protective mecha-
nism, which prevent  
an inadvertent inc-
reasing of the cur-
rent 

if you turn the 

turn 

switch

 first against 

the clockwise di-
rection (lock o-
pens), then in the 
clockwise direction 

wrong operating 
mode 

in the operation 
mode bio feed-
back no current 
flows 

I feel no current 

current intensity is to 
small 

increase the cur-
rent 

 

Содержание Conpreva

Страница 1: ...ck training 13 Electrical stimulation training 15 The combined training Biofeedback electrical stimulation 16 EMG triggered electrical stimulation 18 operation mode pressure 20 preparation 20 set the...

Страница 2: ...triggered electirical stimulation with adjustable pulse width is available The therapeutic use of the Conpreva is based upon a learning strategy whose origin lies within the psychology which aim for r...

Страница 3: ...ase read through this manual completly before you start using the device The device may only be used with the accessories provided by the producer Every use that excesses of this is not allowed Contra...

Страница 4: ...s that emit strong electromagnetic fields to prevent measuring errors In case of allergies towards polymer foam gold and silver please inform your doctor before you start the training Please inform yo...

Страница 5: ...g outside electromagnetic fields could cause measuring errors Clean thoroughly the skin in the places on which you place the electrodes The skin must be perfectly dry and released from hair Use contac...

Страница 6: ...urrent during the exercise Exercise with electrical stimulation uses only channel 1 for measurement so only one emg scale is shown Course of exercise Course of exercise Flowing current in mA and graph...

Страница 7: ...G is overriden correct during maximum tense within the upper range wrong EMG is adjusted too roughly remaining exercise duration current phase battery state EMG range remaining duration of the current...

Страница 8: ...hold it loose and relaxed Electrical stimulation training Accomplish the steps described in preparation Program selection and EMG setting Follow the commands of the voice to relax the musculature The...

Страница 9: ...nimation does not please you you have the possibility to select between different animations Press in addition during the tense or relax phase the Resume button Five further diagrams stand to the sele...

Страница 10: ...increase the current at any time In order to increase the current you must turn first the turn switch thus downward against the clockwise direction If the lock is open you can increase the current The...

Страница 11: ...the EMG tension on the first channel This mode of operation is suitable to measure the pressure in the rectum and the muscle tension of the abdominal muscles at the same time Also the combination of a...

Страница 12: ...l the plug of the sensor cable from the sensor socket and remove the sticking electrodes Keep these for reuse and take care of them as described under maintenance If used remove the headphones Chargin...

Страница 13: ...afety relevant controls in accordance with MedBetreibV 6 prescribed product according with annex 1 of the regulation Attention keep the plugs dry Maintenance Cleaning the sensor Clean the sensor after...

Страница 14: ...tteries or power supply is defect contact the manu factorer on off button doesn t work rechargeable batter ies are fully dis charged charge the batteies Debugging error Reason solution plug of the ada...

Страница 15: ...ters which are important for the therapy e g all times or the frequencies with which the patient has to train Stan dard disease patterns are pre defined The optimal according to scientific knowledge t...

Страница 16: ...grams are available under the menu entry torso To adjust the exercise duration you have to choose Exercise duration within System settings therapy spec parameters The choosen exercise duration does ap...

Страница 17: ...he following way automatically during the training phases the optimal EMG range is constantly selected semiautomatically during the actual EMG setting the patient strongly tense up 5 seconds The optim...

Страница 18: ...ume key Enter the surname of the patient Confirm the input by pressing the resume button Now you can check the entered surname Press the resume button Enter successively first name address date of bir...

Страница 19: ...divided in minutes i e 01 12 means one minute and 12 seconds The scaling of the abscissa orients itself at the maximum EMG value of the respective exercise in V The tense phase is brown represented th...

Страница 20: ...press the turn switch for more then one second Hold it down and turn the turn switch to set the new volume You can change the volume every time excepting the stimulation phase Example Music volume se...

Страница 21: ...rate and Trigger limit with appedant time limit If the trigger automatic is activated the threshold moves down if the patient doesn t achieve the given threshold value The update rate determines the...

Страница 22: ...t for headphones Multilink for connection with DMA or charging device LG PTA med Power supply NiMH recharable batteries size AA 2500 mAh 1 2 V charging device LG PTA 230V 50Hz Technical data Maintenan...

Страница 23: ...8 50 Hz 10A m AC variations IEC 61000 4 11 Voltage Test Level 0 40 and 70 Duration Periods 0 5 5 25 Voltage Interruption 5 sec Security options Prevention of high regulating of the current outside the...

Страница 24: ...es of each exercise are documented so that the execution of the training as well as the adherence to the training defaults can be examined For each patient an allocation of the training data is possib...

Страница 25: ...er authorized by the company Any modifications or repairs performed by unauthorized centers or groups will void this warranty Any malfunction or failure in the product caused by product misuse includi...

Отзывы: