U
T
, >95%
temporary drop)
5s
(On
U
T
, >95% temporary
drop)
PFMF (50/60Hz)
IEC 61000-4-8
400A/m
400A/m/50Hz/60Hz
PFMF should have the
PFMF characteristics of
typical places in
commercial environment
or hospital.
Note: U
T
refers to the AC network voltage before applying test voltage.
Guidance and Manufacturer’s Declaration
- Electromagnetic Immunity
The infusion pump is expected to be used in the following electromagnetic environment. Purchasers
and users should ensure that they will use the product in the following electromagnetic environment:
Immunity
Test
IEC 60601 Test
Level
Compliance
level
Electromagnetic Environment
–
Guidance
RF conduct
GB/T 17626.6
3 V(effective
value)
150 kHz~80 MHz
3 V(effective
value)
Portable and mobile RF communication
device should not be closer to any part of
Hawk-s1 infusion pump than the
recommended isolation distance while
using, including cables. This isolation
distance should be calculated with the
formula corresponding to transmitter
frequency.
Recommended isolation distance