45
pressure range of the gear, if it is low, increase the pressure base value, if it is too low, it will increase
the pressure base value, If it is too high, it will decrease the pressure base value.
Note: It is necessary to observe the patient's physical condition at any time after changing the gear. If
the patient is unwell, immediately stop to infusion and change back to the original gear. The pressure
base value range is adjusted before 0-150.
8.4.16.3 Dedicated tube settings
Select “ON” (open dedicated tube) or “OFF” (close dedicated tube) according to requirement and
press OK key to save.
8.4.16.4 Calibration Operation Settings
Enter the calibration operation setting interface, install the infusion tube as described in 8.1.3, and
prepare the liquid volume/mass measurement device. For details, see 8.3.
8.4.16.5 Maintenance Period
Enter the maintenance cycle setting interface, edit the maintenance time according to the requirement,
0-48 months, press OK key to save after setting.
This function is mainly to remind the user to regularly maintain the machine, set maintenance interval,
when the machine is turned on detects that the distance from the last maintenance has reached the
set time will pop up a reminder box to remind the user to maintain.
8.4.16.6 Restore default setting
After entering “Restore default” setting interface, press
or
to select “cancel” or “confirm”, then
press OK to save setting.
Parameter
Factory defaults
Parameter
Factory defaults
KVO
0.1ml/h
Sound
High
Purge
1200ml/h
Backlight
ON
Bolus
1000ml/h
Key sound
ON
Occlusion
Level 4
Pressure
-
Air Level
OFF
Forget Op
OFF