44
8.4.13 Auto switch off setting
Select “never” or “1-99” minutes according to the requirements and press OK key to save the
parameters
"1-99" minute means that in the non-infusion non-alarm state, Infusion pump automatically shutdown
after "1-99" minutes without operation.
8.4.14 Language setting
After entering “Language” setting interface, press
or
to select required language, then press OK
to save setting.
8.4.15 bed number setting
Enter the bed number setting interface, set the bed number, confirm correct, press OK keyboard to
save.
8.4.16 administrator Settings.
Administrator password: 111111.
8.4.16.1 WIIF configuration and WIFI module (optional)
The infusion pump supports connection to HK-M1000 Hawkmed fluid management system wirelessly,
it works through WIFI module (optional) and router (optional). Steps as below:
WIFI configuration: After entering Settings, select WIFI, set it ON, then active the WIFI function. It will
connect with the WIFI network which connected successfully last time.
WIFI module configuration: Connect with HK-M1000 system wirelessly, send the infusion data, alarms,
reminders etc to HK-M1000 system. For more information please refer to User manual of HK-M1000
Hawkmed fluid management system.
Attention: The configuration must be set by the engineers we admit, or the servicer we appointed.
8.4.16.2 Pressure Base Value Settings
Enter the pressure base value setting interface, edit the pressure base value according to
requirements, confirm and press OK key to save.
The smaller the pressure base value, the higher the blockage alarm sensitivity. This item requires
pressure gauge to calibrate. According to the pressure value of the gear and pressure meter actually
set, adjust this coefficient with reference to table 843-1. If the measured pressure is outside the