Description
Function
Silence key
Press this key to silence the alarm sound. The visible alarm info can’t be
removed by SILENCE key
Power key
Switch on / off the infusion pump.
1.Switch on: in ‘power off’ status, keep pressing this key until LCD screen
displays, and the indicator light is on .
2.Switch off: in ‘switch on’ status, keep pressing this key until hear a sound
“beep” and the indicator light is off.
BOLUS key
In ‘stop’ status, keep pressing ‘BOLUS’ key, the pump starts purging; release
this key, purge stops.
During operation, keep pressing ‘BOLUS’ key, the pump starts
bolus infusion (bolus rate preset by the user). Release the key,
bolus infusion stops and the pump continues infusion at original
rate.
Clear/back key
1. Clear the alarm for both sound and visible info
2. Back to the previous menu
OK key
1. To select the parameter and make it editable
2. Save the setting value
Start key
In ‘stop’ status, press this key to start infusion.
Stop key
Press this key to stop infusion.
Door open key
Under shutdown status, press this key to open the door and switch on the
machine at the same time. This key is invalid during infusing.
AC/Battery
Indicator light
If on, it indicates there’s AC/DC input;
if off, it indicates there’s no AC/DC input.
Indicator light
1. Indicator light indicates operating status/alarms cases.
2. If the IV set is installed correctly and with no air in line, the indicator light
shall be green after the door is closed, which also indicating the pump is
ready for starting infusion.
3. The green indicator light flashes when the infusion is in normal progress.
4. If high-priority alarm occurs during operation, the indicator light shall turn
red and flash.
5. If middle-priority alarm occurs during operation, the indicator light shall
turn yellow and flash.
6. If low-priority alarm occurs during operation, the indicator light shall turn