(3)The parameters of infusion pump should be set or changed by trained and professional staffs.
(4) The Infusion Pump should be placed within 1.2 meters above or below patient’s heart.
(5) If the panel is damaged, please replace it in time to protect the infusion pump from being damaged
by the leaked liquids.
(6) If the infusion pump is not used within specified range of ambient environment, will decrease the
accuracy of infusion and even lead to abnormal operation.
(7) The viscosity and proportion of infusion liquids may affect the accuracy of infusion pump.
(8) The user should calibrate the IV set before using a new brand except the built-in brands in this
infusion pump.
8. Operation Method
In order to ensure infusion accuracy and save the time for calibration, it is suggested to use IV sets of
the following brands: BOON, LX, HD, SG, JR and KDL.
8.1 Infusion Operation
8.1.1 Fix the Infusion Pump and connect it to AC power
Adjust the pole clamp to fix the Infusion Pump properly to a stand and connect it to AC power. The AC
indicator light
(on lower right corner) shall be on.
8.1.2 Switch on/off.
Press Power key
few seconds, the pump will be switched on and do self-test, and shows
Self-testing on the display, it will test: Communication Info, Press sensor, Air bubble sensor,
Dedicated IV set, AC Info and Battery Info. The test result (
√
or
×
) will be displayed on LCD.
Pump self-test information after switching on:
(1) Communication information:
√
indicates CPU normal;
×
indicates CPU abnormal;
(2) Battery information:
√
indicates internal battery normal;
×
indicates internal battery abnormal
(3) AC information:
√
indicates AC connected well;
×
indicates AC fail;
(4) Pressure sensor:
√
indicates normal pressure detection;
×
indicates: abnormal pressure detection;
(5) Air bubble sensor:
√
indicates normal air bubble sensor;
×
indicates abnormal air bubble sensor or
IV set was installed in the pump already;
(6) Dedicated IV set:
√
indicates normal dedicated IV sensor;
×
indicates: abnormal installation of