Operation forgotten
alarm
No operation is made after 2min
when the pump is in
non-operation and non-alarm
status after turning it on.
Press any key to eliminate the alarm
★
Remark: This alarm can be set off, refer
to
8.4.9
of this Manual for details.
Door open alarm
The door of infusion pump is
opened during purging/infusion
process
Press SILENCE key to eliminate alarm
sound; close the door to eliminate alarm.
Near empty
3min before the preset infusion
volume is completed.
Press SILENCE key to eliminate the
alarm
★
Remark: This alarm can be set off, refer
to
8.4.10
of this Manual for details.
Infusion complete
alarm
1. Preset infusion volume is
completed
2. When the cumulative
volumereaches36,000ml
Press SILENCE key to eliminate alarm
sound, press Stop key to eliminate the
alarm and press clear associated key to
clear the cumulative volume.
Air bubble alarm
1 There is air bubble inside
infusion pipe
Press SILENCE to eliminate alarm
sound, open the door to exhaust the air
bubble inside infusion pipe and then
press Start key to restart infusion.
2. Infusion pipe is not installed
properly
Reinstall the infusion pipe correctly
according to
8.1.3.
3. The sensor of infusion pump
is faulted
Please contact the distributor or
manufacturer to repair.
Occlusion alarm
1. The infusion loop is blocked
Press SILENCE key or open the door to
eliminate the alarm.
2. The occlusion sensitivity is
level too high.
Please refer to
8.4.3
of this Manual and
adjust the occlusion pressure alarm
parameter of infusion pump.