47
This option is mainly to display the machine's software version, communication information, battery
information, AC information, and air bubble sensor information.
8.5 Operation Precautions
After the IV set is continuously used for 6 hours, please change the section of IV set that is against
the peristaltic chips, or replace a new one. Meanwhile pay attention to the length of the IV set. Use
extension lines if necessary in case the IV set is stretched out of position when patient turns his
body.
Avoid direct sunlight, high temperature and high humidity.
If the pump work on battery only, please check battery capacity before operation and make sure it
has enough power. Otherwise, recharge the battery fully.
Avoid using the Infusion Pump with problems, which may cause medical accidents and bring harm
to patient’s health and even life.
Only well-trained professionals are permitted to set or adjust infusion parameters.
When infusion at high rate (≥800ml/h), large-sized needle (size 7 or above) should be used,
otherwise it shall influence infusion accuracy.
The Infusion Pump should be placed within 1.2 meters above or below patient’s heart.
The damaged front panel (mask) needs to be replaced in time to prevent leakage.
Infusion Pump works under conditions that exceed the prescribed range may influence infusion
accuracy or even cause malfunction.
The degree of viscosity and ratio of medical liquid may influence infusion accuracy.
The IV set used on this Infusion Pump should get valid Medical Device Registration Certificate.
The Infusion Pump uses ‘Boon’ brand A2 IV set for factory settings. If users use the other brands of
IV set, please calibrate its accuracy on machine before use.
8.6 Contraindications:
No findings so far.
9. Troubleshooting
Failure
Cause Analysis
Solutions