12
G
ENERAL
D
ESCRIPTION
Overtemp Switch
Opens to cut power to the heaters if chamber temperature
reaches (163
°
C
±
4) 325
°
F. Overtemp switch has to be
manually reset.
Valve Shaft Switch
Engaged by the cam on the valve shaft, this switch is
closed only when the control knob is set to PRESSURIZE.
For the exposure phase, the timer will not operate if either
this switch or pressure switch is open. For the purge phase,
this switch must change from the closed state to the open
state. After a 1-minute delay, the timer begins to count
down the preset purge value.
Pressure Switch
Closes when chamber pressure reaches the minimum
exposure pressure of (138 kPa) 20 psi. Opens whenever
chamber pressure is less than that value. For the exposure
phase, the timer will not operate if either this switch or value
switch is open.
Sterilant Level Switch
The sterilant level switch in the reservoir lights the VAPO
indicator when the sterilant level is low. This switch is
preset to actuate when one liter of Vapo-Steril can be
poured into the reservoir. The float level is not adjustable.
Содержание Chemiclave EC5500
Страница 15: ...15 Figure 4 Metering Valve Function PRINCIPLES OF OPERATION ...
Страница 16: ...16 Figure 7 Pressurization PRINCIPLES OF OPERATION Figure 6 Ready Load Figure 5 Warm Up ...
Страница 17: ...17 Figure 8 EC5500 EC6000 Operation PRINCIPLES OF OPERATION ...
Страница 18: ...18 Figure 11 End Unload Figure 10 Purge Figure 9 Exposure PRINCIPLES OF OPERATION ...
Страница 20: ...20 Figure 13 Piping Schematic PRINCIPLES OF OPERATION ...
Страница 21: ...21 Figure 14 Cycle Phase Diagram PRINCIPLES OF OPERATION ...
Страница 22: ...22 Door Gasket Figure 15 Exploded View Top Cover and Door PRINCIPLES OF OPERATION ...
Страница 32: ...32 MAINTENANCE AND SERVICING Figure 17 Metering Valve need new drawing ...
Страница 39: ...39 MAINTENANCE AND SERVICING Figure 22 Power Supply Printed Circuit Board ...
Страница 56: ...56 ...