47
Publication 5419-013 REV1.0
Pump Chain Commands
GANG
Sets or displays the syringe count. The Gang command is not valid
in Independent and Reciprocating Condition. In Twin Condition,
Gang may be set to either 1 or 2.
If an attempt is made to set the Gang count to higher than the
allowed value, the response will be:
<lf>[<addr>]Range error:<sp>{requested gang setting}
<lf>[<addr>]<sp><sp><sp>Syringe count out of range of <min>
to <max>.
Command format (Twin Condition):
gang[<sp>{syringe count}
Query response to gang:
<lf>[<addr>]# syringes<cr>
<lf>[<addr>]<prompt>
Query response to gang<sp>{syringe count}:
<lf>[<addr>]<prompt>
Example:
gang<sp>ab<sp>1
SVOLUME
Sets or displays syringe volume.
Command format (Independent Condition):
svolume<sp>{a|b|ab}[<sp>{syringe volume} {ul|ml}]
Query response to svolume<sp>{a|b|ab}
<lf>[<addr>]<axis># xl<cr> (displays one line per requested axis)
<lf>[<addr>]<prompt> (displayed after last axis line)
Query response to svolume
<sp>{a|b|ab}<sp>{syringevolume {ul|ml}}
<lf>[<addr>]<prompt>
Command format (Reciprocating or Twin Condition):
svolume[<sp>{syringe volume} {ul|ml}]
Query response to svolume
<lf>[<addr>]#
xl<cr>
<lf>[<addr>]<prompt>
Query response to svolume<sp>{syringevolume {ul|ml}}
<lf>[<addr>]<prompt>
Examples:
svolume<sp>ab<sp>5<sp>ul
svolume<sp>10<sp>ml
DIAMETER
Sets or displays the syringe diameter in mm.
Command format (Independent Condition):
diameter<sp>{a|b|ab}[<sp>{syringe diameter}]
Query response to diameter<sp>{a|b|ab}:
<lf>[<addr>]<axis><sp># (displays one line per requested axis)
<lf>[<addr>]<prompt>
Query response to diameter<sp>{a|b|ab}<sp>[{syringe
diameter}]:
<lf>[<addr>]<prompt>
Command format (Reciprocating or Twin Condition):
diameter[<sp>{syringe diameter}]
Query response to diameter:
<lf>[<addr>]#
mm
<lf>[<addr>]<prompt>
Query response to diameter<sp>{syringe diameter}:
<lf>[<addr>]<prompt>
Example:
diameter<sp>a<sp>4.85