Twin Condition Setup
Twin Condition Syringe Select Screen
(Showing Gang x1 and 2.5 ml Syringe Selection)
3) RATE SELECT – Select the Infuse, Withdraw, Infuse/Withdraw
or Withdraw/Infuse flow rates. The Rate Selection will vary
depending on the mode selected. The Infuse, Withdraw,
Infuse/Withdraw or Withdraw/Infuse rate entry is the same as
covered in the Independent Condition Rate Select section.
a. Infuse Mode Rate Select – Select the infuse rate for P1.
The same rate will be automatically set for Pump Channel
2 (P2) infuse rate.
b. Withdraw Mode Rate Select – Select the withdraw rate
for P1. The same rate will automatically set for the P2
withdraw rate.
c. Infuse/Withdraw Mode Rate Select – Select the Infuse rate
and Withdraw rate for P1. The user can set different rates
for Infuse and Withdraw. The same rates will automatically
be set for P2.
d. Withdraw/Infuse Mode Rate Select – Select the Withdraw
rate and Infuse rate. The user can set different rates for
Infuse and Withdraw. The same rates will automatically be
set for P2.
NOTE: Please see 4 below for details on the impact of the Ganging
feature on the Rate Select setting.
4) TARGET SELECT – Select the Target Volume or Time depending on
the Mode selected. The Infuse, Withdraw, Infuse/Withdraw or
Withdraw/Infuse Target Select entry is the same as covered in the
Independent Condition Target Select section.
NOTE: Please see below for details on the impact of the Ganging
feature on the Rate and Target Select settings.
a. Ganging Feature RATE AND TARGET SELECT – The Ganging
feature described above impacts the Rate Select setting. When
the user selects the x2 Gang feature in the Twin Condition
Select Syringe table, the flow rate and target ranges listed
are double the allowed flow rate and target ranges of a
single syringe.
This requires the user to connect the output of both syringes
together using the same tubing size and length to produce the
correct flow rates. The following are examples using the x1
and x2 Gang feature.
Example x1 Ganging: The user requires the proper setting for
the following application, Mode Select setting of Infuse, using a
2.5 ml syringe (inner diameter 7.285 mm). The user wants both
syringes to infuse at the same rate and volume. The syringe
outputs will not be connected together. In this case, the user
will select the Gang x1 toggle selection located in the top right
hand corner of the Select Syringe menu along with the proper
2.5 ml syringe.
The x1 selection allows both syringes to operate as two identical
individual syringes with the same mode, syringe, flow rate and target
volume or time. The x2 selection doubles the flow and volume.
Note: This requires the user to connect the output of both syringes
together using the same tubing size and length.