53
• The short name of the selected language (3 chars)
NSLx
Requests the number of logged samples (4 chars).
The command parameter (1 char):
• P ‑ request for pH range
• M ‑ request for mV and Rel mV ranges
• I ‑ request for ISE range
LODPxxx
Requests the xxxth pH record logged data.
LODMxxx
Requests the xxxth mV/Rel mV record logged data.
LODIxxx
Requests the xxxth ISE record logged data (
HI98191
).
LODPALL
Requests all pH Log on demand.
LODMALL
Requests all mV/Rel mV Log on demand.
LODIALL
Requests all ISE Log on demand (
HI98191
).
The answer string for each record contains:
• The logged mode (2 chars):
• 00 ‑ pH range (0.001 resolution)
• 01 ‑ pH range (0.01 resolution)
• 02 ‑ pH range (0.1 resolution)
• 03 ‑ mV range
• 04 ‑ Rel mV range
• 05 ‑ ISE range
• Reading status (1 char): R, O, U
• Calculated reading, with sign and decimal point and exponent
(11 chars) ‑ for pH, Rel mV and ISE range
• Temperature reading, with sign and two decimal points (7 chars)
• mV reading status (1 char): R, O, U
• The mV reading, with sign and decimal point (7 chars)
• The logged time,
yymmddhhmmss
(12 chars)
• The calibration slope, with sign and decimal point (7 chars) ‑ not available
for Rel mV range
• The calibration offset, with sign and decimal point (7 chars) ‑ not available
for ISE
• Temperature probe presence (1 char)
PC INTERF
ACE