Wrong slope
The calibration cannot be confirmed.
This message is displayed if slope is out of the accepted range.
Slope under accepted value (30 % default slope).
Slope over accepted value (130 % default slope).
Wrong old slope
An inconsistency between new and previous (old) calibration is detected. Clear old calibration
parameters and proceed calibration from the current point. The instrument will keep all confirmed
values during current calibration.
The instrument will display “‑‑‑‑” on the primary LCD if is not calibrated or after all calibrations are
cleared.
If
“Clear”
is pressed during the first calibration point, the instrument returns to measurement
mode.
Notes: Press T functional key or MODE to select temperature value to be changed if the
temperature probe is not connected.
ISE range is not temperature compensated.
ISE CALIBRA
TION HI98191