SPECIFICA
TIONS
3. SPECIFICATIONS
pH specifications
EC specifications
TDS specifications
*Note: pH range will be reduced to actual probe specification
** Note: 1000 μS/cm = 500 ppm with 0.5 TDS Factor
= 700 ppm with 0.7 TDS Factor
Range
0.00 to 14.00 pH, 0.0 to 14.0 pH*
Resolution
0.01 pH, 0.1 pH
Accuracy (@25 ºC/77 ºF)
±0.05 pH, ±0.1 pH
Calibration
one or two‑point calibration (using pH 4.01, 7.01, 10.01
buffers) using auto buffer recognition
one‑point calibration using quick calibration solution
Temperature compensation
automatic: 0.0 to 60.0 °C; 32.0 to 140.0 °F
Range
0.00 to 10.00 mS/cm
Resolution
0.01 mS/cm
Accuracy (@25 ºC/77 ºF)
±0.10 mS/cm (0.00 to 5.00 mS/cm)
±0.20 mS/cm (5.00 to 10.00 mS/cm)
Calibration
one‑point at 1.41 mS/cm or 5.00 mS/cm using auto
standard recognition.
one‑point calibration using quick calibration solution
Temperature compensation
(not user selectable)
automatic, with
b
= 1.9%/°C
Range
0 to 5000 ppm (0.5 TDS Factor)**
0 to 7000 ppm (0.7 TDS Factor)**
Resolution
10 ppm
Accuracy (@25 ºC/77 ºF)
±2% FS
Calibration
through EC calibration
TDS factor
0.5 (500 ppm) or 0.7 (700 ppm)