When the reading is stable and close to the selected buffer, the
Confirm
key will become available.
Press
Confirm
to accept and store the calibration point 4.01 or 10.01pH.
One-point calibration
To perform a one‑point calibration only, press
Escape
to calibrate in 7.01 pH, or, the
Monitor
will
automatically save one calibration point for 4.01 and 10.01 pH. It will store the calibration infor‑
mation and return to MENU.
6.1.3. EC CALIBRATION
Select
EC
from MENU and press
CAL
to enter the EC calibration function.
The following functions are available in EC Calibration mode:
•
Clear:
Press to clear the last calibration from the
Monitor
. After confirming, the “Calibration
cleared” message is displayed.
•
Confirm:
Press to accept the current calibration point which is only available if the measurement
is stable and within the limits for the selected EC standard solution.
•
Escape:
Press to exit calibration and return to MENU.
While in
EC Calibration
mode, the display will show the current EC reading, the current temperature
reading and the current standard:
The
Monitor
stores the two‑point calibration information and returns to MENU mode.