Note: The GLP history will show "Calibration cleared" message with the date.
Immerse the probe in 1.41 mS/cm or 5.00 mS/cm calibration solution.
From the MENU screen with
EC
highlighted, press the
CAL
key to begin the calibration process.
EC standard value will be autorecognized by the
Monitor
.
When the reading is stable and close to the selected standard, the
Confirm
key will become available.
Press
Confirm
to accept and store the calibration point.
Press
Escape
to exit calibration without saving the calibration.
OPERA
TIONAL GUIDE
The calibration can be erased at any time by entering the calibration and pressing
Clear
. A warning
message appears and confirmation for deleting the calibration is requested. Press
Accept
to confirm
or
Cancel
to exit and return to calibration screen.