81
• Insert the cuvet with the reacted sample
(# 2) into the instrument. Ensure
that the notch on the cap is positioned
securely into the groove.
• Press READ DIRECT and “SIP” will blink during measurement.
• The instrument directly displays concentration in µg/L of hydra-
zine on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines
# 2
Hydrazine