108
SPECIFICATIONS
Range
0 to 15000 mg/L
Resolution
10 mg/L (1 mg/L x 10)
Precision
Standard deviation ±220 @ 10000 mg/L
Typical EMC
±1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 610 nm
Method
Adaptation of the
EPA method 410.4
. Oxidizable
organic compounds reduce the dichromate ion
(orange) to the Chromium (III) ion (green). The
amount of Chromium (III) formed is determined.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93754C
Liquid Reagent
25 vials
REAGENT SETS AND ACCESSORIES
C 9800-01
Hanna Reactor (115 VAC)
C 9800-02
Hanna Reactor (230 VAC)
HI 93754C-25
Reagents for up to 24 tests
HI 740216
Test tube cooling rack (25 holes)
HI 740217
Laboratory bench safety shield
For other accessories see page 141.
MEASUREMENT PROCEDURE
Before starting to use the reagent kit it is important to read carefully all
the instructions and the Health & Safety Data Sheet
(HSDS). Pay particular attention to all warnings, cau-
tions and notes. Failure to do so may result in serious
injury to the operator.
•
Choose a homogeneous sample. Samples containing settleable
solids need to be homogenized with a blender.
CHEMICAL OXYGEN DEMAND
CHEMICAL OXYGEN DEMAND
CHEMICAL OXYGEN DEMAND
CHEMICAL OXYGEN DEMAND
CHEMICAL OXYGEN DEMAND
HIGH RANGE
HIGH RANGE
HIGH RANGE
HIGH RANGE
HIGH RANGE
COD HR