76
TOTAL HARDNESS LOW RANGE
TOTAL HARDNESS LOW RANGE
TOTAL HARDNESS LOW RANGE
TOTAL HARDNESS LOW RANGE
TOTAL HARDNESS LOW RANGE
SPECIFICATIONS
Range
0 to 250 mg/L
Resolution
1 mg/L from 0 to 100 mg/L
5 mg/L from 100 to 250 mg/L
Accuracy
±5 mg/L ±4% of reading
Typical EMC
±5 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 466 nm
Method
Adaptation of the
EPA method 130.1
.
The reac-
tion between calcium, magnesium and reagents
causes a red-violet tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93735
A-LR
Hardness Indicator reagent LR 9.5 mL
HI 93735
B
-0
Hardness Buffer reagent
2 drops
HI 93735
C
-0
Fixing reagent
1 packet
REAGENT SETS
HI 93735-00
Reagents for 100 tests Low Range (0 to 250 mg/L)
HI 93735-0
Reagents for 100 tests (0 to 750 mg/L)
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Total Hardness LR
on the secondary LCD by pressing PROGRAM
W
and
V
.
• Using the syringe, add exactly 0.5 mL of
unreacted sample to the cuvet.
• With the plastic pipette fill the
cuvet up to the 10 mL mark
adding HI 93735A-LR Indicator
reagent.
Total Hardness LR