SPECIFICATIONS
Range
0 to 500 PCU (Platinum Cobalt Units)
Resolution
1 PCU
Accuracy
±10 PCU ±5% of reading
Typical EMC
± 1 PCU
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 420 nm
Method
Adaptation of the
Standard Methods for the
Examination of Water and Wastewater, 18
th
edition
, Colorimetric Platinum Cobalt method.
REQUIRED ACCESSORIES
0.45 µm membrane for true color measurement.
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Color of Water on
the secondary LCD by pressing PROGRAM
W
and
V
.
• Fill one cuvet up to the mark with deionized
water and replace the cap. This is the blank.
• Place the blank (# 1) into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
• Press ZERO and "SIP" will blink on the display.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
• Remove the blank.
COLOR OF WATER
COLOR OF WATER
COLOR OF WATER
COLOR OF WATER
COLOR OF WATER
Color of Water
10 mL
# 1
# 1