54
COPPER LOW RANGE
COPPER LOW RANGE
COPPER LOW RANGE
COPPER LOW RANGE
COPPER LOW RANGE
SPECIFICATIONS
Range
0 to 1000 µg/L
Resolution
1 µg/L
Accuracy
±10 µg/L ±5% of reading
Typical EMC
±1 µg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 575 nm
Method
Adaptation of the
EPA method.
The reaction
between copper and the bicinchoninate reagent
causes a purple tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93747-0
Bicinchoninate
1 packet
REAGENT SETS
HI 93747-01
Reagents for 100 tests
HI 93747-03
Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Copper LR on the
secondary LCD by pressing PROGRAM
W
and
V
.
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed and
ready for measurement.
Copper LR
10 mL