52
Copper HR
COPPER HIGH RANGE
COPPER HIGH RANGE
COPPER HIGH RANGE
COPPER HIGH RANGE
COPPER HIGH RANGE
SPECIFICATIONS
Range
0.00 to 5.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.02 mg/L ±4% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 575 nm
Method
Adaptation of the
EPA method.
The reaction
between copper and the bicinchoninate reagent
causes a purple tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93702-0
Bicinchoninate
1 packet
REAGENT SETS
HI 93702-01
Reagents for 100 tests
HI 93702-03
Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Copper HR on the
secondary LCD by pressing PROGRAM
W
and
V
.
• Fill the cuvet with 10 mL of unreacted
sample (up to the mark) and replace the
cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
10 ml
Содержание C 200 Series
Страница 146: ...146 USER NOTES USER NOTES USER NOTES USER NOTES USER NOTES ...
Страница 147: ...147 ...
Страница 148: ...148 W W W H A N N A I N S T C O M MANC200R4 08 03 ...