Änderungen vorbehalten / Subject to change without notice
5 6
Forms:
Command and query.
Parameters: None.
9.3
*ESR?
Description: Standard Event Status Register Query. Reading this register clears it. This
register is defined in IEEE-488 11.5.1.2.
Forms:
Query only.
Parameters: None.
9.4
*IDN?
Description: Identification Query. The response from *IDN? identifies a unique
instrument. The response consists of a single string containing four
separate fields, comma delimited.
The fields are: <manufacturer>,<model>,<serial number >,<version>
Where
<version> is divided as: <interface version/hardware version/firmware
version>
Forms:
Query only.
Parameters: None.
9.5
*OPC
*OPC?
Description: Operation Complete Command/Query. The *OPC command is intended to
cause the instrument to generate a flag indicating when all pending
commands have been processed. The *OPC? is intended to place an ASCII
‘1’ in the instrument’s output buffer if all pending commands are complete.
Forms:
Command and query.
Parameters: None.
9.6
*RST
Description: Reset Command. The Reset command performs a single interface reset
and it sets the control values to the factory default values.
Forms:
Command only.
Parameters: None.
9.7
*SRE
*SRE?
Description: Sets or returns the Service Request Enable Register. This register is
defined in IEEE-488 11.3.2.
Forms:
Command and query.
Parameters: None.
9.8
*STB?
Description: Read Status Byte Query. This register is defined in IEEE-488 11.36.
Forms:
Query only.
Parameters: None.
9.9
*TST?
Description: Self-test Query. A value of zero indicates the self-test has been completed
without errors detected.
Forms:
Query only.
Parameters: None.