71

Command Reference

CALL 1

The CALL 1 command determines the bridge settings and 
ensures that the subsequent command (CROP or CRSH) 
performs an open or short compensation for all 69 test fre-

quencies. It is necessary to first send CROP or CRSH be

-

fore the compensation is performed.

CROP

The CROP command performs an open compensation. 
The HM8118 automatically reports immediately if a com-
pensation was successful (0) or if it failed (-1).

CRSH

The CRSH command performs a short compensation. The 
HM8118 automatically reports immediately if a compensa-
tion was successful (0) or if it failed (-1).

XALL?

The XALL? query queries the measurement values of the 
main measurement value display, the secondary measure-
ment value display and the number of sorting bins. The 
measurement values are issued separated by a comma. If 
the binning interface is not activated / not integrated or if 
the current measurement is invalid, the sorting bin value 
99 issued.

XMAJ?

The  XMAJ? query queries the measurement value of the 
main measurement value display. If the measurement va-
lue display is set to percentage deviation and if the nomi-
nal measurement value is „0“, an error message will be 
issued.

XMIN? 

The XMIN? query queries the measurement value of the 
secondary measurement value display. If the measurement 
value display is set to percentage deviation and if the no-
minal measurement value is „0“, an error message will be 
issued.

XDLT?

The XDLT? query queries the absolute deviation between 
the measurement value and the nominal measurement va-
lue (see also PREL command). If the automatic measure-
ment mode (AUTO) is selected, an error message is 
issued.

XDMT?

The  XDMT? query queries the relative deviation between 
the measurement value and the nominal measurement va-
lue (see also PREL command). If the nominal measurement 
value is set to „0“ or if the automatic measurement mode 
(AUTO) is selected, an error message is issued.

10.3  Command List Binning Interface

(only with integrated binning interface HO118)

XBIN?

The XBIN? query queries the number of sorting bins for the 

current measurement. If the binning interface is not swit-
ched on / not activated or if the current measurement is in-
valid, the sorting bin value 99 issued.

BBUZ(?) i

The BBUZ command activates (i=1) or deactivates (i=0) the 
alarm function of the binning interface. The BBUZ? query 
queries the current status of the alarm function.

BCLR

The BCLR command deletes the nominal values and limits 
for all sorting bins. It also deactivates the binning interface.

BING(?) {i}

The BING command locks (i=0) and enables (i=1) the bin-
ning. If no sorting bin is opened or if the measurement 
mode „AUTO“ is selected for the HM8118 , an error mes-
sage is issued.

BLIH j,(?) {x}

 

The BLIH command sets the maximum limit (i = 0) of a 

sorting bin j to x % between 0 and 7. The  BLIH? query 

queries the maximum limit (i = 0) of the sorting bin.

BLIL j,(?) {x}

The BLIL command sets the lower limit (i = 1) for a sorting 

bin j x % between 0 and 7. The lower limit must be less 

than or equal to the upper limit. If no lower limit has been 
set, the HM8118 applies the negative value of the upper li-
mit as lower limit. The query BLIL? queries the lower limit (i 
= 1) of the sorting bin.

BNOM i,(?) {x}

  

The BNOM command set sets the nominal value of the 
sorting bin i to the value x. The value i can be anywhere 
between 0 and 8 (sorting bin 8 is the QDR sorting bin for 
failures). If no nominal value has been set for the sorting 
bin, the HM8118 applies the nominal value of the subse-
quent lowest numbered sorting bin with a nominal va-
lue of unequal 0 (multiple sorting bins can have the identi-
cal nominal value without having a value entered for each 
sorting bin). The lowest numbered active sorting bin must 
have a selected nominal value. The sorting bin 0 must al-
ways be set for the binning to work. The query BNOM? 
queries the nominal value of the sorting bin.

Содержание HM8118

Страница 1: ...HM8118 Programmable LCR Bridge User Manual...

Страница 2: ...not leave the premises All signal connec tions must be shielded e g coax such as RG58 U With signal generators double shielded cables are manda tory It is especially important to establish good ground...

Страница 3: ...rement Range 51 6 5 Circuit Type 52 7 Instrument Functions 52 7 1 SETUP Menu 52 7 1 1 Measurement Frequency FRQ 52 7 1 2 Voltage LEV 53 7 1 3 Preload Bias Current BIAS 53 7 1 4 Measurement Range RNG 5...

Страница 4: ...fety reasons the instrument may only be operated with authorized safety sockets The power cord must be plugged in before signal circuits may be connected Never use the product if the power cable is da...

Страница 5: ...n the back panel has to be unplugged The product may only be opened by authorized and qualified personnel Prior to working on the product or before the product is opened it must be disconnected from t...

Страница 6: ...Cells and batteries must be recycled and kept separate from residual waste Cells and batteries must be recy cled and kept separate from residual waste Recharge able batteries and normal batteries tha...

Страница 7: ...ng the parallel equivalent circuit RANGE 17 AUTO HOLD Activating the automatic measurement range LED lights up or the range HOLD function 18 UP Range up 19 DOWN Range down Connectors 20 L CUR BNC sock...

Страница 8: ...38 L Q 1 Selection of the measurement function Induc tance and Quality factor Q resp entering numeric value 1 39 Z 7 Selection of the measurement function Im pedance Z and Phase angle resp entering nu...

Страница 9: ...be very low at this setting The smaller the loss angle the more the real world com ponents will come close to the ideal An ideal inductor has a loss angle of zero degrees An ideal capacitor also has a...

Страница 10: ...esistors will be measured using the parallel equivalent circuit 4 First Time Ope ration 4 1 Connecting the instrument Prior to connecting the instrument to the mains check whether the mains voltage co...

Страница 11: ...result of multiple possibly overlapping individual capacities inductances and impedances This can easily cause some misunderstandings especially with coils magnetic field eddy currents hysteresis etc...

Страница 12: ...d by the dissipation factor DF The equivalent series resistance ESR is frequency depen dent according to the formula ESR Rs D Cs where Omega 2 f circular frequency represents Traditionally the inducta...

Страница 13: ...only be displayed with an integrated binning interface HO118 Use the keys L R 2 36 C D 3 34 C R 4 32 R Q 5 30 to select the sub menus associated with the menu functions Depending on the function you...

Страница 14: ...elative measurement value deviation in percent Press the SELECT key 3 to use the SETUP and the set ting DEV_M for the main measurement value display and DEV_S for the secondary measurement value displ...

Страница 15: ...nts the instrument from constantly switching the measurement range if the measurement va lue is close to the switching threshold of a measurement range The following table shows the switching threshol...

Страница 16: ...be better suited as equivalent circuit diagram for the measurement circuit In the auto matic mode the bridge selects the serial equivalent circuit diagram for impedances below 1k and the parallel equ...

Страница 17: ...m i e after switching off the bias current it is required to wait for the coils to discharge before the component is disconnected from the measuring instrument During the discharge Please wait is show...

Страница 18: ...e RNG The measurement range can be selected automatically or manually If the measurement range is changed the inter nal measurement circuit replacement circuit will be mo dified and internal relays wi...

Страница 19: ...rigger event measurement has been triggered or the selected measurement function has been selected incorrectly lue via keypad confirm the entry by pressing the ENTER key or by pressing the knob again...

Страница 20: ...g HZ184 are used If the DUT ex hibits impedances of more than 25k at frequencies of more than 100kHz is is also recommended to use the GUARD function REF_S In the SETUP menu use the arrow keys 7 to s...

Страница 21: ...ithout components connected The short compensation is possible for impedances of up to 15 and resistances of up to 10 LOAD A compensation with adjustment compensation of a known load impedance is suit...

Страница 22: ...enter a value After en tering the value via keypad confirm the entry by pressing the ENTER key or by pressing the knob again The LOADM or LOADS function is not necessary for Hameg ac cessories In this...

Страница 23: ...FRQ function to choose the line frequency press the knob 6 editing mode and select the desired line frequency 50Hz 60Hz via knob Press the knob again to confirm the selection 7 3 6 Instrument Informat...

Страница 24: ...nt panel During a measure ment do not touch the component directly with hands or indi rectly with objects as this may distort the measurement results Always remove measurement accessories such as test...

Страница 25: ...rate via 4 wire con nection with LCR bridge HM8118 Measurable components Resistances coils or capacitators Frequency range 20 Hz to 200 kHz Test lead length approximately 35 cm Connectors BNC sockets...

Страница 26: ...he unknown state of the prema gnetized core The component to be measured is depen dent on the frequency as well as on the applied measure ment voltage The measuring instrument determines the values fo...

Страница 27: ...de termining the leakage inductance requires the secondary transformer side to be short circuited see fig 8 11 If the secondary side is short circuited the measured values of the primary side correspo...

Страница 28: ...easurement se tup are implemented it becomes necessary to perform a new compensation For more information about the OPEN SHORT compensation see chapter 7 2 8 5 Sorting Components with Option HO118 Bin...

Страница 29: ...nge for the sorting bins 0 5 and only the secondary parameters exceed the limit for sorting bin 6 7 General Fai lure BIN This sorting bin output is activated if the sorting does not apply to one of th...

Страница 30: ...sponds to TTL level and triggers with falling slopes It is protected against voltages of up to 15 volts For more information on the binning interface in the con text of the PIN and jumper assignment r...

Страница 31: ...E 488 controller control unit of a IEEE 488 bus system Any PC equipped with the corresponding plug in card can function as IEEE 488 controller If a IEC 625 cable must be used an appropriate plug in ad...

Страница 32: ...ort parallel processing of commands measuring instrument and if the instrument is ready to per form another measurement WAI The WAI command is a synchronization command that stops every subsequent com...

Страница 33: ...easurement circuit By de fault the automatic circuit type i 2 is selected Use i 0 The constant voltage CST V function must be switched on for measurements with bias current or external preload The err...

Страница 34: ...ative measurement value deviation REF_M for the main measurement value display if the measurement value de viation of the main measurement value display was previ ously activated via OUTP 1 or OUTP 2...

Страница 35: ...o 0 or if the automatic measurement mode AUTO is selected an error message is issued 10 3 Command List Binning Interface only with integrated binning interface HO118 XBIN The XBIN query queries the nu...

Страница 36: ...to 180 999 99 to 999 99 M 1 H to 100H N 0 95 to 500 Measurement conditions and functions Test frequency 20Hz to 200kHz 69 steps Frequency accuracy 100ppm AC test signal level 50mVrms to 1 5Vrms Resol...

Отзывы: