First-Time Operation

shown in the right panel and the essential parameters in 
the left panel of the display. The four BNC sockets located 
on the front panel can be connected to the component to 
be measured with the appropriate measuring accessories. 
Additionally, it is also possible to connect the measuring 
instrument via ground socket on the front panel 

43

 with 

ground potential. The socket is suitable for a banana plug 
with a 4 mm diameter.

If there are undefined messages on the display or if the in

-

strument fails to react to operation of its controls turn it 
off, wait a minute and turn it on again in order to trigger a 
reset operation. If the display remains unchanged or ope-

ration impossible, turn it off and take it to a qualified ser

-

vice point (see Safety Instructions).

4.3  Line frequency

Prior to first measurements, the line frequency setting 

must be set to the applied line frequency, 50 or 60 Hz. If 
the line frequency is not set properly, depending on the 
measurement range and the line frequency value, insta-
bilities may occur e.g. on the display. In order to set the 
line frequency press the SELECT button 

3

, use the SYST 

menu for accessing MAINS FRQ, use the knob 

6

 for se-

lecting the correct value.

4.4  Measurement Principle

The LCR meter HM8118 is not a traditional Wien, Maxwell 
or Thomson measurement bridge. Rather, when connec-
ting a test object, the impedance |Z| and the correspon-

ding phase angle Θ (phase between current and voltage) 
are always determined (see fig. 4.3). These measurement 

values are frequency dependent and will be determined 
by means of an AC test signal (which can be set manually 
between 50mV and 1.5V). The test signal is induced in the 
test item. This distinguishes a LCR bridge from a multime-
ter (DC measurement). Based on the measurement prin-
ciple, the measured impedance is always essential. Based 
on the impedance (X axis) and the phase angel (angle), the 
instrument is able to determine the missing value of the 
Y axis. This means that it is not the DC component that is 
being measured but rather the AC value. The issued values 
are calculated digitally. This measurement of impedance 
and phase angle is subject to a certain measurement inac-
curacy which will be described on the following pages.

In general, the HM8118 bridge can only determine the 
ESR, ESC or ESL (= Equivalent Series Resistance / Capa-
city / Inductivity) according to the equivalent circuit dia-

The front panel ground connector and the ground contact of the 
trigger input are directly connected to the mains safety ground 
potential through the line cord. The outer contacts of the front 
panel BNC connectors 

20

 – 

23

 (as well as the shields of any co-

axial cables attached) are connected to the GUARD potential 
which has no connection to the safety ground! No external vol-
tages may be applied to the BNC connectors! The rear panel in-
terfaces 

47

 and 

48

 are galvanically isolated (no connection to 

ground)!

gram of the component and is primarily used to measure 
individual components. If a circuit with multiple compo-
nents is connected to the bridge, the instrument will al-
ways determine the ESR, ESC or ESL of the entire circuit 
/ component group. This can potentially skew the measu-
rement result. The connected component / circuit is assu-
med to be the „Black Box“. These values are available for 
each component; however, please keep in mind that these 
always describe the result of multiple, possibly overlapping 
individual capacities, inductances and impedances. This 
can easily cause some misunderstandings especially with 

coils (magnetic field, eddy currents, hysteresis, etc.)

Fig. 4.4 shows the link between capacity C

s

 (or resistance 

R

s

) and various test voltages that can be selected with the 

bridge (0.2V

eff

 to 1.5V

eff

). As can be seen in the figure, the 

measurement values of C

s

 or R

s

 are highly dependent on 

the selected test voltage. Point A shows the test point of 
the instrument during the measurement of a single com-
ponent, point B shows the test point during the measu-
rement of a component group (in this case two capaci-
ties connected in parallel). In contrast to test point A, with 
point B the bridge switches the measurement range due 
to the impedance of the entire component group. As a re-
sult, the measurement results for point A and point B are 
different.

The LCR bridge HM8118 is primarily intended to determine pas-
sive components. Therefore, it is not possible to determine test 
objects which are externally supplied with power.

applied on the red terminal.

The bias voltage works only

when the instrument on capacitance measurement 
mode.

Measuring function selection 

The desired test function is selected by push buttons (12) 
and (14). The push button (12) gives access to the main
parameter (R, L or C), The push button (14) allows a 
secondary parameter measurement  (Q/D, impedance or
phase). 
In order to measure D parameter the instrument needs at
first to be set to capacitance measurement mode, on the
other way, Q parameter will be displayed.

Auto-measurement function 

The HM8018-2 is able to automatically determine the 
component type in most cases. 3 different automatisms 
exists: the automatic impedance range selection (see the
section « Auto-ranging»), the automatic mode 
(series/parallel) selection

(see the section « 

passive 

components »), and the automatic function selection. These 
three automatisms are simultaneously activated when the
instrument is set in automatic mode with the

RANGE

AUTO

 key (7). Then the user can change function or mode 

that disables their respective automatism. The manual range
selection disables the three automatisms. 
When the instrument is on automatic mode the function
choice depends on the impedance module, phase angle as
well as the quality factor .The diagram below shows the 
choice made by the instrument. 

|Z| = 1000 

Ω

Q = 500

D = 0,002

Q = 500

D = 0,002

D = 500

Q = 0,002

D = Q = 1

D = Q = 1

R

jX

D = 500

Q = 0,002

Calculation functions 

Apart from displaying normal values as resistance, inductance or
capacitance, the HM8018-2 can display relative deviations and 
percentages. It is not possible to use these calculation modes for 
other functions than the three previous values. The deviations
and percentages are displayed in relation to the two stored values
A and B. 

The procedure to obtain relative measurement is as follows: 
1)

Connect the component corresponding to the reference 
value. 

2)

Store the value (memory A) by pressing on the 

STORE

key,

then press the 

A

key. 

3)  Press on the 

A

key. The indicator 

-A

lights up and the 

display shows the value (

Measure

– A). 

A direct percentage measurement is possible, it is only to use the
÷

B

key instead of the 

–A

key in the previous procedure. Then the

instrument displays the value 100*

Measure

/B in %.

To obtain a deviation in % proceed as follows: 

1)

Connect the component corresponding to the reference 
value. 

Imaginary 

Axis

below -- 45° = C

above 45° = L

Real 

Axis

ph

ase

 an

gle

Q = Quality factor
D = Loss tangent

D = 1 / Q
Q = 1 / D = 1 / tan delta
(delta = Opposite angle of the phase angle)

Fig. 4.3: Measurement principle

Содержание HM8118

Страница 1: ...HM8118 Programmable LCR Bridge User Manual...

Страница 2: ...not leave the premises All signal connec tions must be shielded e g coax such as RG58 U With signal generators double shielded cables are manda tory It is especially important to establish good ground...

Страница 3: ...rement Range 51 6 5 Circuit Type 52 7 Instrument Functions 52 7 1 SETUP Menu 52 7 1 1 Measurement Frequency FRQ 52 7 1 2 Voltage LEV 53 7 1 3 Preload Bias Current BIAS 53 7 1 4 Measurement Range RNG 5...

Страница 4: ...fety reasons the instrument may only be operated with authorized safety sockets The power cord must be plugged in before signal circuits may be connected Never use the product if the power cable is da...

Страница 5: ...n the back panel has to be unplugged The product may only be opened by authorized and qualified personnel Prior to working on the product or before the product is opened it must be disconnected from t...

Страница 6: ...Cells and batteries must be recycled and kept separate from residual waste Cells and batteries must be recy cled and kept separate from residual waste Recharge able batteries and normal batteries tha...

Страница 7: ...ng the parallel equivalent circuit RANGE 17 AUTO HOLD Activating the automatic measurement range LED lights up or the range HOLD function 18 UP Range up 19 DOWN Range down Connectors 20 L CUR BNC sock...

Страница 8: ...38 L Q 1 Selection of the measurement function Induc tance and Quality factor Q resp entering numeric value 1 39 Z 7 Selection of the measurement function Im pedance Z and Phase angle resp entering nu...

Страница 9: ...be very low at this setting The smaller the loss angle the more the real world com ponents will come close to the ideal An ideal inductor has a loss angle of zero degrees An ideal capacitor also has a...

Страница 10: ...esistors will be measured using the parallel equivalent circuit 4 First Time Ope ration 4 1 Connecting the instrument Prior to connecting the instrument to the mains check whether the mains voltage co...

Страница 11: ...result of multiple possibly overlapping individual capacities inductances and impedances This can easily cause some misunderstandings especially with coils magnetic field eddy currents hysteresis etc...

Страница 12: ...d by the dissipation factor DF The equivalent series resistance ESR is frequency depen dent according to the formula ESR Rs D Cs where Omega 2 f circular frequency represents Traditionally the inducta...

Страница 13: ...only be displayed with an integrated binning interface HO118 Use the keys L R 2 36 C D 3 34 C R 4 32 R Q 5 30 to select the sub menus associated with the menu functions Depending on the function you...

Страница 14: ...elative measurement value deviation in percent Press the SELECT key 3 to use the SETUP and the set ting DEV_M for the main measurement value display and DEV_S for the secondary measurement value displ...

Страница 15: ...nts the instrument from constantly switching the measurement range if the measurement va lue is close to the switching threshold of a measurement range The following table shows the switching threshol...

Страница 16: ...be better suited as equivalent circuit diagram for the measurement circuit In the auto matic mode the bridge selects the serial equivalent circuit diagram for impedances below 1k and the parallel equ...

Страница 17: ...m i e after switching off the bias current it is required to wait for the coils to discharge before the component is disconnected from the measuring instrument During the discharge Please wait is show...

Страница 18: ...e RNG The measurement range can be selected automatically or manually If the measurement range is changed the inter nal measurement circuit replacement circuit will be mo dified and internal relays wi...

Страница 19: ...rigger event measurement has been triggered or the selected measurement function has been selected incorrectly lue via keypad confirm the entry by pressing the ENTER key or by pressing the knob again...

Страница 20: ...g HZ184 are used If the DUT ex hibits impedances of more than 25k at frequencies of more than 100kHz is is also recommended to use the GUARD function REF_S In the SETUP menu use the arrow keys 7 to s...

Страница 21: ...ithout components connected The short compensation is possible for impedances of up to 15 and resistances of up to 10 LOAD A compensation with adjustment compensation of a known load impedance is suit...

Страница 22: ...enter a value After en tering the value via keypad confirm the entry by pressing the ENTER key or by pressing the knob again The LOADM or LOADS function is not necessary for Hameg ac cessories In this...

Страница 23: ...FRQ function to choose the line frequency press the knob 6 editing mode and select the desired line frequency 50Hz 60Hz via knob Press the knob again to confirm the selection 7 3 6 Instrument Informat...

Страница 24: ...nt panel During a measure ment do not touch the component directly with hands or indi rectly with objects as this may distort the measurement results Always remove measurement accessories such as test...

Страница 25: ...rate via 4 wire con nection with LCR bridge HM8118 Measurable components Resistances coils or capacitators Frequency range 20 Hz to 200 kHz Test lead length approximately 35 cm Connectors BNC sockets...

Страница 26: ...he unknown state of the prema gnetized core The component to be measured is depen dent on the frequency as well as on the applied measure ment voltage The measuring instrument determines the values fo...

Страница 27: ...de termining the leakage inductance requires the secondary transformer side to be short circuited see fig 8 11 If the secondary side is short circuited the measured values of the primary side correspo...

Страница 28: ...easurement se tup are implemented it becomes necessary to perform a new compensation For more information about the OPEN SHORT compensation see chapter 7 2 8 5 Sorting Components with Option HO118 Bin...

Страница 29: ...nge for the sorting bins 0 5 and only the secondary parameters exceed the limit for sorting bin 6 7 General Fai lure BIN This sorting bin output is activated if the sorting does not apply to one of th...

Страница 30: ...sponds to TTL level and triggers with falling slopes It is protected against voltages of up to 15 volts For more information on the binning interface in the con text of the PIN and jumper assignment r...

Страница 31: ...E 488 controller control unit of a IEEE 488 bus system Any PC equipped with the corresponding plug in card can function as IEEE 488 controller If a IEC 625 cable must be used an appropriate plug in ad...

Страница 32: ...ort parallel processing of commands measuring instrument and if the instrument is ready to per form another measurement WAI The WAI command is a synchronization command that stops every subsequent com...

Страница 33: ...easurement circuit By de fault the automatic circuit type i 2 is selected Use i 0 The constant voltage CST V function must be switched on for measurements with bias current or external preload The err...

Страница 34: ...ative measurement value deviation REF_M for the main measurement value display if the measurement value de viation of the main measurement value display was previ ously activated via OUTP 1 or OUTP 2...

Страница 35: ...o 0 or if the automatic measurement mode AUTO is selected an error message is issued 10 3 Command List Binning Interface only with integrated binning interface HO118 XBIN The XBIN query queries the nu...

Страница 36: ...to 180 999 99 to 999 99 M 1 H to 100H N 0 95 to 500 Measurement conditions and functions Test frequency 20Hz to 200kHz 69 steps Frequency accuracy 100ppm AC test signal level 50mVrms to 1 5Vrms Resol...

Отзывы: