69

Command Reference

measurement values to be used for the calculation of the 
average. The AVGM? query queries the status of the calcu-
lation of the average.

NAVG(?) {i}

If you use the AVGM command to set the calculation of 
the average to i=1, you can use the NAVG command to set 
the number of measurement values to be used for the cal-
culation of the average anywhere between 2 and 99. 
The NAVG? query queries the number of measurement va-
lues to be used for the calculation of the average.

VBIA(?) {x}

The VBIA command sets one internal DC preload any-
where between 0V and 5V. This command returns an error 
message (ERROR) if the HM8118 is not set to a measure-

ment mode C-D, C-R, R-X or Z-Θ which would be suitable 

for a preload. Use the BIAS 1 (= internal) command to ac-
tivate the preload which was previously activated via VBIA 
and to show it on the display. The VBIA? query queries the 
current value of the applied DC preload.

IBIA(?) {x}

The IBIA command defines the DC bias current between

0.001A and 0.200A. This command returns an error mes-
sage (ERROR) if the HM8118 is not set to perform an in-
ductance measurement or a transformer measurement (L-

Q, L-R, N-Θ or M). Use the BIAS 1 (= internal) command 

to activate the bias current which was previously activa-
ted via IBIA and to show it on the display. The IBIA? query 
queries the current DC bias current.

BIAS(?) {i}

The BIAS command activates or deactivates the DC pre-

load or DC bias current defined in the HM8118. Use i=0 

to deactivate the DC preload that is selected via VBIA, or 
deactivate the DC bias current that is selected via IBIA. 
Use i=1 to activate the internal BIAS and to show the va-
lue that was previously selected via VBIA or IBIA on the 
display. Use i=2 to select the external BIAS which is only 
possible with a DC preload. The internal BIAS preload can 
only be selected if the instrument is set to the appropri-
ate measurement function (see VBIA command). The in-
ternal BIAS bias current can only be selected if the instru-
ment is set to the appropriate measurement function (see 
IBIA command). The external BIAS function behaves cor-
respondingly. The BIAS? query queries the current BIAS 
status.

CIRC(?) {i}

Use the CIRC command to select the circuit type of the 
equivalent circuit diagram (measurement circuit). By de-
fault, the automatic circuit type (i=2) is selected. Use i=0 

The constant voltage (CST V function) must be switched on for 
measurements with bias current or external preload.

The error message "DCR too high“ indicates that the resistance 
of the connected DUT is too high for the selected bias current. In 
this case, the bias current cannot be activated.

to set the series circuit, use i=1 to set the parallel circuit 
of the equivalent circuit diagram. The CIRC? query queries 
the current status of the equivalent circuit diagram setting.

CONV(?) {i}

The CONV command activates (i=1) or deactivates (i=0) 
the constant voltage (function CST V). The CONV? query 
queries the current status of the constant voltage.

FREQ(?) {x}

Use the FREQ command to select the measurement fre-
quency in Hz. The 69 available measurement frequency in-
tervals are as follows:

Measurement Frequencies

20Hz

90Hz

500Hz

2.5kHz

12kHz

72kHz

24Hz

100Hz

600Hz

3.0kHz

15kHz

75kHz

25Hz

120Hz

720Hz

3.6kHz

18kHz

80kHz

30Hz

150Hz

750Hz

4.0kHz

20kHz

90kHz

36Hz

180Hz

800Hz

4.5kHz

24kHz

100kHz

40Hz

200Hz

900Hz

5.0kHz

25kHz

120kHz

45Hz

240Hz

1.0kHz

6.0kHz

30kHz

150kHz

50Hz

250Hz

1.2kHz

7.2kHz

36kHz

180kHz

60Hz

300Hz

1.5kHz

7.5kHz

40kHz

200kHz

72Hz

360Hz

1.8kHz

8.0kHz

45kHz

75Hz

400Hz

2.0kHz

9.0kHz

50kHz

80Hz

450Hz

2.4kHz

10kHz

60kHz

The FREQ? query queries the set measurement frequency 
in Hz.

MMOD(?) {i}

Use the MMOD command to select the trigger type. Use 
i=0 to select the continuous trigger, i.e. a new measure-
ment will automatically be performed upon completion of 
the previous measurement. Use i=1 to select the manual 
trigger (TGM). In this case, a measurement will be perfor-
med only after the *TRG command was sent. Use i=2 to 
select the external trigger (TGE). A measurement is perfor-
med when a rising slope is applied to the external trigger 
input (TTL level +5V). The MMOD? query queries the cur-
rent status of the triggering.

*TRG / STRT

Use the *TRG or STRT command to start a measurement
if the manual trigger mode was previously selected (see
MMOD).

RATE(?) {i}

Use the RATE command to set the measurement speed 
(SPD function) in the increments FAST (i=0), MED (i=1) 
or SLOW (i=2). The number of measurements for a con-
tinuous triggering (CONT) is approximately 1.5 per se-

If the measuring instrument shows a blank screen (i.e. lines "- - 
-“) without measurement values, no trigger event / measurement 
has been triggered or the selected measurement function has 
been selected incorrectly.

Содержание HM8118

Страница 1: ...HM8118 Programmable LCR Bridge User Manual...

Страница 2: ...not leave the premises All signal connec tions must be shielded e g coax such as RG58 U With signal generators double shielded cables are manda tory It is especially important to establish good ground...

Страница 3: ...rement Range 51 6 5 Circuit Type 52 7 Instrument Functions 52 7 1 SETUP Menu 52 7 1 1 Measurement Frequency FRQ 52 7 1 2 Voltage LEV 53 7 1 3 Preload Bias Current BIAS 53 7 1 4 Measurement Range RNG 5...

Страница 4: ...fety reasons the instrument may only be operated with authorized safety sockets The power cord must be plugged in before signal circuits may be connected Never use the product if the power cable is da...

Страница 5: ...n the back panel has to be unplugged The product may only be opened by authorized and qualified personnel Prior to working on the product or before the product is opened it must be disconnected from t...

Страница 6: ...Cells and batteries must be recycled and kept separate from residual waste Cells and batteries must be recy cled and kept separate from residual waste Recharge able batteries and normal batteries tha...

Страница 7: ...ng the parallel equivalent circuit RANGE 17 AUTO HOLD Activating the automatic measurement range LED lights up or the range HOLD function 18 UP Range up 19 DOWN Range down Connectors 20 L CUR BNC sock...

Страница 8: ...38 L Q 1 Selection of the measurement function Induc tance and Quality factor Q resp entering numeric value 1 39 Z 7 Selection of the measurement function Im pedance Z and Phase angle resp entering nu...

Страница 9: ...be very low at this setting The smaller the loss angle the more the real world com ponents will come close to the ideal An ideal inductor has a loss angle of zero degrees An ideal capacitor also has a...

Страница 10: ...esistors will be measured using the parallel equivalent circuit 4 First Time Ope ration 4 1 Connecting the instrument Prior to connecting the instrument to the mains check whether the mains voltage co...

Страница 11: ...result of multiple possibly overlapping individual capacities inductances and impedances This can easily cause some misunderstandings especially with coils magnetic field eddy currents hysteresis etc...

Страница 12: ...d by the dissipation factor DF The equivalent series resistance ESR is frequency depen dent according to the formula ESR Rs D Cs where Omega 2 f circular frequency represents Traditionally the inducta...

Страница 13: ...only be displayed with an integrated binning interface HO118 Use the keys L R 2 36 C D 3 34 C R 4 32 R Q 5 30 to select the sub menus associated with the menu functions Depending on the function you...

Страница 14: ...elative measurement value deviation in percent Press the SELECT key 3 to use the SETUP and the set ting DEV_M for the main measurement value display and DEV_S for the secondary measurement value displ...

Страница 15: ...nts the instrument from constantly switching the measurement range if the measurement va lue is close to the switching threshold of a measurement range The following table shows the switching threshol...

Страница 16: ...be better suited as equivalent circuit diagram for the measurement circuit In the auto matic mode the bridge selects the serial equivalent circuit diagram for impedances below 1k and the parallel equ...

Страница 17: ...m i e after switching off the bias current it is required to wait for the coils to discharge before the component is disconnected from the measuring instrument During the discharge Please wait is show...

Страница 18: ...e RNG The measurement range can be selected automatically or manually If the measurement range is changed the inter nal measurement circuit replacement circuit will be mo dified and internal relays wi...

Страница 19: ...rigger event measurement has been triggered or the selected measurement function has been selected incorrectly lue via keypad confirm the entry by pressing the ENTER key or by pressing the knob again...

Страница 20: ...g HZ184 are used If the DUT ex hibits impedances of more than 25k at frequencies of more than 100kHz is is also recommended to use the GUARD function REF_S In the SETUP menu use the arrow keys 7 to s...

Страница 21: ...ithout components connected The short compensation is possible for impedances of up to 15 and resistances of up to 10 LOAD A compensation with adjustment compensation of a known load impedance is suit...

Страница 22: ...enter a value After en tering the value via keypad confirm the entry by pressing the ENTER key or by pressing the knob again The LOADM or LOADS function is not necessary for Hameg ac cessories In this...

Страница 23: ...FRQ function to choose the line frequency press the knob 6 editing mode and select the desired line frequency 50Hz 60Hz via knob Press the knob again to confirm the selection 7 3 6 Instrument Informat...

Страница 24: ...nt panel During a measure ment do not touch the component directly with hands or indi rectly with objects as this may distort the measurement results Always remove measurement accessories such as test...

Страница 25: ...rate via 4 wire con nection with LCR bridge HM8118 Measurable components Resistances coils or capacitators Frequency range 20 Hz to 200 kHz Test lead length approximately 35 cm Connectors BNC sockets...

Страница 26: ...he unknown state of the prema gnetized core The component to be measured is depen dent on the frequency as well as on the applied measure ment voltage The measuring instrument determines the values fo...

Страница 27: ...de termining the leakage inductance requires the secondary transformer side to be short circuited see fig 8 11 If the secondary side is short circuited the measured values of the primary side correspo...

Страница 28: ...easurement se tup are implemented it becomes necessary to perform a new compensation For more information about the OPEN SHORT compensation see chapter 7 2 8 5 Sorting Components with Option HO118 Bin...

Страница 29: ...nge for the sorting bins 0 5 and only the secondary parameters exceed the limit for sorting bin 6 7 General Fai lure BIN This sorting bin output is activated if the sorting does not apply to one of th...

Страница 30: ...sponds to TTL level and triggers with falling slopes It is protected against voltages of up to 15 volts For more information on the binning interface in the con text of the PIN and jumper assignment r...

Страница 31: ...E 488 controller control unit of a IEEE 488 bus system Any PC equipped with the corresponding plug in card can function as IEEE 488 controller If a IEC 625 cable must be used an appropriate plug in ad...

Страница 32: ...ort parallel processing of commands measuring instrument and if the instrument is ready to per form another measurement WAI The WAI command is a synchronization command that stops every subsequent com...

Страница 33: ...easurement circuit By de fault the automatic circuit type i 2 is selected Use i 0 The constant voltage CST V function must be switched on for measurements with bias current or external preload The err...

Страница 34: ...ative measurement value deviation REF_M for the main measurement value display if the measurement value de viation of the main measurement value display was previ ously activated via OUTP 1 or OUTP 2...

Страница 35: ...o 0 or if the automatic measurement mode AUTO is selected an error message is issued 10 3 Command List Binning Interface only with integrated binning interface HO118 XBIN The XBIN query queries the nu...

Страница 36: ...to 180 999 99 to 999 99 M 1 H to 100H N 0 95 to 500 Measurement conditions and functions Test frequency 20Hz to 200kHz 69 steps Frequency accuracy 100ppm AC test signal level 50mVrms to 1 5Vrms Resol...

Отзывы: