Hameg HM8118 Скачать руководство пользователя страница 15

51

Setting of Parameters

surement value (reference value) is displayed. The measu-
rement value deviation is shown in the appropriate units 
(Ohm, Henry, etc.).

6.3  Reference Value (REF_M, REF_S)

The menu function REF_M or REF_S enables the user 
to enter a reference value which will be used as a basis 

for the measurement result ∆ % or ∆ ABS. Press the SE

-

LECT key 

3

 to use the SETUP menu function and the set-

ting REF_M (for the main measurement value display) and 
REF_S (for the secondary measurement value display) to 
enter a reference value each. The applicable units will be 
selected automatically depending on the selected measu-
rement function for the main measurement value display 

(H, F, Ω or S) or for the secondary measurement value dis

-

play (Ω, S or °). You can enter a reference value numerically 
with up to five digits after the decimal point. Alternatively, 

you can press the TRIG key 

27

 to perform a measurement, 

and the resulting measurement value will be adopted as 
reference value.

6.4  Selecting the Measurement Range

The measurement range can be selected automatically or 
manually. In some cases, it is useful to lock the automa-
tic measurement range function as it can take a complete 
measurement cycle to determine the appropriate measure-
ment range. This can also be useful when switching similar 
components. The bridge HM8118 automatically switches 
to the measurement range 6 and subsequently back to the 
adequate measurement range if a component has been 
connected to the instrument. If the automatic measure-
ment range function has been locked and the impedance 
of a component equals more than 100 times the nominal 
value of the measurement range, the bridge will display an 
OVERRANGE measurement error. In this case, it is neces-
sary to select a suitable measurement range for the mea-
surement. Press the AUTO/HOLD key 

17

 to switch bet-

ween the automatic and the manual measurement range 
selection.

6.4.1   Automatic range selection (AUTO)

If the automatic measurement range function is activated, 
the bridge automatically selects the most suitable mea-
surement range for an exact measurement in accordance 
with the connected component. The instrument will switch 
to the next measurement range level below if the measure-

ment value is smaller than 22.5% of the selected measure

-

ment range or 90% higher than the end value of the mea

-

surement range. An integrated switching hysteresis of ap-

proximately 10% prevents the instrument from constantly 

switching the measurement range if the measurement va-
lue is close to the switching threshold of a measurement 
range. The following table shows the switching thresholds 
for switching the measurement range (if the constant vol-
tage CST V is switched off):

Measurement Range

Component Impedance

1 to 2

Z > 3.00 Ω

2 to 3

Z > 100.00 Ω

3 to 4

Z > 1.60 kΩ

4 to 5

Z > 25.00 kΩ

5 to 6

Z > 1.00 MΩ

2 to 1

Z < 2.70 Ω

3 to 2

Z < 90.00

 

Ω

4 to 3

Z < 1.44

 

5 to 4

Z < 22.50

 

6 to 5

Z < 900.00

 

6.4.2  Manual Measurement Range Selection

The bridge HM8118 includes 6 measurement ranges (1–6). 
The measurement ranges can be preselected manually or 
automatically. The following table indicates the source re-
sistance and the impedance of the connected component 

for each measurement range. The specified ranges are im

-

pedance ranges, not resistance ranges. Capacitators or in-
ductances are frequency-dependent components.

Measure-

ment range

Source 

Impedance

Component 

Impedance

1

25.0 Ω

10.0 µΩ bis 3.0 Ω

2

25.0 Ω

3.0 Ω bis 100.0 Ω

3

400.0 Ω

100.0 Ω bis 1.6 kΩ

4

6.4 kΩ

1.6 kΩ bis 25.0 kΩ

5

100.0 kΩ

25.0 kΩ bis 2.0 MΩ

6

100.0 kΩ

2.0 MΩ bis 100.0 MΩ

Additionally, the impedance of capacitators is inversely 
proportionate to the frequency. Therefore, larger capaci-
tators will be measured in the lower impedance measure-
ment ranges. Consequently, the measurement range for 
any given component may change as the measurement 
frequency changes. If you wish to measure multiple si-
milar components, it is possible to shorten the measure-
ment time by using the AUTO/HOLD 

17

 key to switch from 

the automatic measurement range selection to the manual 
measurement range selection with the DUT (= Device Un-
der Test) connected. The AUTO/HOLD key will no longer 
be illuminated. It is recommended to primarily use the ma-

During the measurement of an inductance in the AUTO mode, it 
may occur that the HM8118 is constantly changing the measure-
ment range. This is based on the fact that the source impedance 
is dependent on the selected measurement range so that after 
switching the measurement range, the newly measured value is 
outside the range of the 10% hysteresis. In this case, it is recom-
mended to use the manual measurement range selection.

The LCR bridge HM8118 does not create a 50Ω system. Instead, 

it changes its internal resistance dependent on measurement 
function and measurement range. Every cable shows losses and 
distorts the original measurement result because of inductive and 
capacitive properties (particularly because of its length).
The input impedance changes dependent on the selected 
measurement range and the connect load impedance between 

25Ω and 100kΩ.

Содержание HM8118

Страница 1: ...HM8118 Programmable LCR Bridge User Manual...

Страница 2: ...not leave the premises All signal connec tions must be shielded e g coax such as RG58 U With signal generators double shielded cables are manda tory It is especially important to establish good ground...

Страница 3: ...rement Range 51 6 5 Circuit Type 52 7 Instrument Functions 52 7 1 SETUP Menu 52 7 1 1 Measurement Frequency FRQ 52 7 1 2 Voltage LEV 53 7 1 3 Preload Bias Current BIAS 53 7 1 4 Measurement Range RNG 5...

Страница 4: ...fety reasons the instrument may only be operated with authorized safety sockets The power cord must be plugged in before signal circuits may be connected Never use the product if the power cable is da...

Страница 5: ...n the back panel has to be unplugged The product may only be opened by authorized and qualified personnel Prior to working on the product or before the product is opened it must be disconnected from t...

Страница 6: ...Cells and batteries must be recycled and kept separate from residual waste Cells and batteries must be recy cled and kept separate from residual waste Recharge able batteries and normal batteries tha...

Страница 7: ...ng the parallel equivalent circuit RANGE 17 AUTO HOLD Activating the automatic measurement range LED lights up or the range HOLD function 18 UP Range up 19 DOWN Range down Connectors 20 L CUR BNC sock...

Страница 8: ...38 L Q 1 Selection of the measurement function Induc tance and Quality factor Q resp entering numeric value 1 39 Z 7 Selection of the measurement function Im pedance Z and Phase angle resp entering nu...

Страница 9: ...be very low at this setting The smaller the loss angle the more the real world com ponents will come close to the ideal An ideal inductor has a loss angle of zero degrees An ideal capacitor also has a...

Страница 10: ...esistors will be measured using the parallel equivalent circuit 4 First Time Ope ration 4 1 Connecting the instrument Prior to connecting the instrument to the mains check whether the mains voltage co...

Страница 11: ...result of multiple possibly overlapping individual capacities inductances and impedances This can easily cause some misunderstandings especially with coils magnetic field eddy currents hysteresis etc...

Страница 12: ...d by the dissipation factor DF The equivalent series resistance ESR is frequency depen dent according to the formula ESR Rs D Cs where Omega 2 f circular frequency represents Traditionally the inducta...

Страница 13: ...only be displayed with an integrated binning interface HO118 Use the keys L R 2 36 C D 3 34 C R 4 32 R Q 5 30 to select the sub menus associated with the menu functions Depending on the function you...

Страница 14: ...elative measurement value deviation in percent Press the SELECT key 3 to use the SETUP and the set ting DEV_M for the main measurement value display and DEV_S for the secondary measurement value displ...

Страница 15: ...nts the instrument from constantly switching the measurement range if the measurement va lue is close to the switching threshold of a measurement range The following table shows the switching threshol...

Страница 16: ...be better suited as equivalent circuit diagram for the measurement circuit In the auto matic mode the bridge selects the serial equivalent circuit diagram for impedances below 1k and the parallel equ...

Страница 17: ...m i e after switching off the bias current it is required to wait for the coils to discharge before the component is disconnected from the measuring instrument During the discharge Please wait is show...

Страница 18: ...e RNG The measurement range can be selected automatically or manually If the measurement range is changed the inter nal measurement circuit replacement circuit will be mo dified and internal relays wi...

Страница 19: ...rigger event measurement has been triggered or the selected measurement function has been selected incorrectly lue via keypad confirm the entry by pressing the ENTER key or by pressing the knob again...

Страница 20: ...g HZ184 are used If the DUT ex hibits impedances of more than 25k at frequencies of more than 100kHz is is also recommended to use the GUARD function REF_S In the SETUP menu use the arrow keys 7 to s...

Страница 21: ...ithout components connected The short compensation is possible for impedances of up to 15 and resistances of up to 10 LOAD A compensation with adjustment compensation of a known load impedance is suit...

Страница 22: ...enter a value After en tering the value via keypad confirm the entry by pressing the ENTER key or by pressing the knob again The LOADM or LOADS function is not necessary for Hameg ac cessories In this...

Страница 23: ...FRQ function to choose the line frequency press the knob 6 editing mode and select the desired line frequency 50Hz 60Hz via knob Press the knob again to confirm the selection 7 3 6 Instrument Informat...

Страница 24: ...nt panel During a measure ment do not touch the component directly with hands or indi rectly with objects as this may distort the measurement results Always remove measurement accessories such as test...

Страница 25: ...rate via 4 wire con nection with LCR bridge HM8118 Measurable components Resistances coils or capacitators Frequency range 20 Hz to 200 kHz Test lead length approximately 35 cm Connectors BNC sockets...

Страница 26: ...he unknown state of the prema gnetized core The component to be measured is depen dent on the frequency as well as on the applied measure ment voltage The measuring instrument determines the values fo...

Страница 27: ...de termining the leakage inductance requires the secondary transformer side to be short circuited see fig 8 11 If the secondary side is short circuited the measured values of the primary side correspo...

Страница 28: ...easurement se tup are implemented it becomes necessary to perform a new compensation For more information about the OPEN SHORT compensation see chapter 7 2 8 5 Sorting Components with Option HO118 Bin...

Страница 29: ...nge for the sorting bins 0 5 and only the secondary parameters exceed the limit for sorting bin 6 7 General Fai lure BIN This sorting bin output is activated if the sorting does not apply to one of th...

Страница 30: ...sponds to TTL level and triggers with falling slopes It is protected against voltages of up to 15 volts For more information on the binning interface in the con text of the PIN and jumper assignment r...

Страница 31: ...E 488 controller control unit of a IEEE 488 bus system Any PC equipped with the corresponding plug in card can function as IEEE 488 controller If a IEC 625 cable must be used an appropriate plug in ad...

Страница 32: ...ort parallel processing of commands measuring instrument and if the instrument is ready to per form another measurement WAI The WAI command is a synchronization command that stops every subsequent com...

Страница 33: ...easurement circuit By de fault the automatic circuit type i 2 is selected Use i 0 The constant voltage CST V function must be switched on for measurements with bias current or external preload The err...

Страница 34: ...ative measurement value deviation REF_M for the main measurement value display if the measurement value de viation of the main measurement value display was previ ously activated via OUTP 1 or OUTP 2...

Страница 35: ...o 0 or if the automatic measurement mode AUTO is selected an error message is issued 10 3 Command List Binning Interface only with integrated binning interface HO118 XBIN The XBIN query queries the nu...

Страница 36: ...to 180 999 99 to 999 99 M 1 H to 100H N 0 95 to 500 Measurement conditions and functions Test frequency 20Hz to 200kHz 69 steps Frequency accuracy 100ppm AC test signal level 50mVrms to 1 5Vrms Resol...

Отзывы: