Instrument Functions

20Hz and 200kHz. In the CORR menu, use the arrow keys 

 

7

 to select the FRQ function to choose the mea-

surement frequency, press the knob 

6

 (editing mode) and 

select the desired measurement frequency via knob. Press 

the knob again to confirm the selection.

7.2.4  FUNC Function

You can use the FUNC function to select the measurement 
function for the load impedance LOADM and LOADS. You 
can choose from the following functions:

Ls-Q, 

Lp-Q, 

Ls-Rs, 

Lp-Rp, 

Cs-D, 

Cp-D, 

Cs-Rs, 

Cp-Rp, 

Rs-Q, 

Rp-Q, 

Z-Θ,

Y-Θ,

R-X 

G-B

In the CORR menu, use the arrow keys 

 

7

 to se-

lect the FUNC function to choose the measurement func-
tion, press the knob 

6

 (editing mode) and select the desi-

red function via knob. Press the knob again to confirm the 

selection.

7.2.5  Correction Factors LOADM / LOADS

You can use the LOADM function (main measurement va-
lue display) to save a reference value for the load impe-
dance LOAD in the reference memory LOADM. Depending 
on the parameter FUNC H, you can choose one of the fol-
lowing as unit for the measurement value: mH, µH, nH, F, 

mF, µF, nF, pF, Ω, mΩ, kΩ, MΩ, or S, kS, mS, µS, nS, pS. 

In the CORR menu, use the arrow keys 

 

7

 to se-

lect the LOADM function to set the reference value, press 
the knob 

6

 (editing mode) and select the desired refe-

rence value via knob. By pressing the knob again, you can 
activate the manual value input via numeric keypad. A va-
lue input window will be opened. You can use the numeric 
keys to enter a value. After entering the value via keypad, 

confirm the entry by pressing the ENTER key or by pres

-

sing the knob again.

You can use the LOADS function (secondary measurement 
value display) to save a reference value for the load impe-
dance LOAD in the reference memory LOADS. Depending 
on the parameter FUNC, you can choose one of the fol-

lowing as unit for the measurement value: Ω, mΩ, kΩ, MΩ, 

S, kS, mS, µS, nS, pS or °. In the CORR menu, use the ar-
row keys 

 

7

 to select the LOADS function to set 

the reference value, press the knob 

6

 (editing mode) and 

select the desired reference value via knob. By pressing 
the knob again, you can activate the manual value input 
via numeric keypad. A value input window will be opened. 
You can use the numeric keys to enter a value. After en-

tering the value via keypad, confirm the entry by pressing 

the ENTER key or by pressing the knob again.

The LOADM or LOADS function is not necessary for Hameg ac-
cessories. In this case, the regular OPEN /SHORT compensation 

is sufficient.

You can use the parameters LOADM and LOADS if it is dif-

ficult to align a connect measurement adapter or if it is 

connected to the bridge via long test leads. In this case, a 
complete open or short circuit compensation is not pos-
sible because the bridge cannot compensate the actual 
equivalent circuit diagram of the measurement adapter 
with a simple equivalent circuit. This places the bridge in a 
state that cannot be compensated. The user can compen-
sate the measurement error by means of a known impe-
dance with a given frequency. 

If the compensation with a known load (LOAD) is activa-
ted, the bridge corrects the measurement value of the con-
nected impedance in relation to three impedances:

Short circuit impedance,

Idle time impedance

Load impedance

It is possible to use up to 5 different reference values for 
the load impedance which can be selected by means of 
NUM parameter. An impedance always corresponds to a 
group of parameters: a number, a frequency, a function 
and naturally the known parameters of the impedance.

After the compensation with adjustment (LOAD), the impe-
dance is connected to the measured impedance to mea-
sure with the load impedance correction. Correcting with 
a load impedance is most effective if the load impedance 
is near the measured impedance. If the compensation with 
adjustment (LOAD) is switched on (parameter LOAD set to 
„ON“), the load impedance correction is automatically ac-
tivated when the set measurement frequency is equal to 
the measurement frequency of the load impedance LOAD 
which is saved for the load impedance corrections within 
the 5 parameter groups. Therefore, it is important that the 
5 parameter groups have different frequencies for the load 
impedance correction.

7.3  Menu Function SYST

7.3.1 CONTRAST Function

You can use the CONTRAST function to set the display 
contrast from 35 to 55. In the SYST menu, use the ar-

For a compensation with adjustment, a value is entered in both 
reference memories (LOADM and LOADS) (e.g. for a real re-
sistance for LOADM the resistance value and the value "0" for 
LOADS).

Fig. 7.9: Menu function SYST display

Содержание HM8118

Страница 1: ...HM8118 Programmable LCR Bridge User Manual...

Страница 2: ...not leave the premises All signal connec tions must be shielded e g coax such as RG58 U With signal generators double shielded cables are manda tory It is especially important to establish good ground...

Страница 3: ...rement Range 51 6 5 Circuit Type 52 7 Instrument Functions 52 7 1 SETUP Menu 52 7 1 1 Measurement Frequency FRQ 52 7 1 2 Voltage LEV 53 7 1 3 Preload Bias Current BIAS 53 7 1 4 Measurement Range RNG 5...

Страница 4: ...fety reasons the instrument may only be operated with authorized safety sockets The power cord must be plugged in before signal circuits may be connected Never use the product if the power cable is da...

Страница 5: ...n the back panel has to be unplugged The product may only be opened by authorized and qualified personnel Prior to working on the product or before the product is opened it must be disconnected from t...

Страница 6: ...Cells and batteries must be recycled and kept separate from residual waste Cells and batteries must be recy cled and kept separate from residual waste Recharge able batteries and normal batteries tha...

Страница 7: ...ng the parallel equivalent circuit RANGE 17 AUTO HOLD Activating the automatic measurement range LED lights up or the range HOLD function 18 UP Range up 19 DOWN Range down Connectors 20 L CUR BNC sock...

Страница 8: ...38 L Q 1 Selection of the measurement function Induc tance and Quality factor Q resp entering numeric value 1 39 Z 7 Selection of the measurement function Im pedance Z and Phase angle resp entering nu...

Страница 9: ...be very low at this setting The smaller the loss angle the more the real world com ponents will come close to the ideal An ideal inductor has a loss angle of zero degrees An ideal capacitor also has a...

Страница 10: ...esistors will be measured using the parallel equivalent circuit 4 First Time Ope ration 4 1 Connecting the instrument Prior to connecting the instrument to the mains check whether the mains voltage co...

Страница 11: ...result of multiple possibly overlapping individual capacities inductances and impedances This can easily cause some misunderstandings especially with coils magnetic field eddy currents hysteresis etc...

Страница 12: ...d by the dissipation factor DF The equivalent series resistance ESR is frequency depen dent according to the formula ESR Rs D Cs where Omega 2 f circular frequency represents Traditionally the inducta...

Страница 13: ...only be displayed with an integrated binning interface HO118 Use the keys L R 2 36 C D 3 34 C R 4 32 R Q 5 30 to select the sub menus associated with the menu functions Depending on the function you...

Страница 14: ...elative measurement value deviation in percent Press the SELECT key 3 to use the SETUP and the set ting DEV_M for the main measurement value display and DEV_S for the secondary measurement value displ...

Страница 15: ...nts the instrument from constantly switching the measurement range if the measurement va lue is close to the switching threshold of a measurement range The following table shows the switching threshol...

Страница 16: ...be better suited as equivalent circuit diagram for the measurement circuit In the auto matic mode the bridge selects the serial equivalent circuit diagram for impedances below 1k and the parallel equ...

Страница 17: ...m i e after switching off the bias current it is required to wait for the coils to discharge before the component is disconnected from the measuring instrument During the discharge Please wait is show...

Страница 18: ...e RNG The measurement range can be selected automatically or manually If the measurement range is changed the inter nal measurement circuit replacement circuit will be mo dified and internal relays wi...

Страница 19: ...rigger event measurement has been triggered or the selected measurement function has been selected incorrectly lue via keypad confirm the entry by pressing the ENTER key or by pressing the knob again...

Страница 20: ...g HZ184 are used If the DUT ex hibits impedances of more than 25k at frequencies of more than 100kHz is is also recommended to use the GUARD function REF_S In the SETUP menu use the arrow keys 7 to s...

Страница 21: ...ithout components connected The short compensation is possible for impedances of up to 15 and resistances of up to 10 LOAD A compensation with adjustment compensation of a known load impedance is suit...

Страница 22: ...enter a value After en tering the value via keypad confirm the entry by pressing the ENTER key or by pressing the knob again The LOADM or LOADS function is not necessary for Hameg ac cessories In this...

Страница 23: ...FRQ function to choose the line frequency press the knob 6 editing mode and select the desired line frequency 50Hz 60Hz via knob Press the knob again to confirm the selection 7 3 6 Instrument Informat...

Страница 24: ...nt panel During a measure ment do not touch the component directly with hands or indi rectly with objects as this may distort the measurement results Always remove measurement accessories such as test...

Страница 25: ...rate via 4 wire con nection with LCR bridge HM8118 Measurable components Resistances coils or capacitators Frequency range 20 Hz to 200 kHz Test lead length approximately 35 cm Connectors BNC sockets...

Страница 26: ...he unknown state of the prema gnetized core The component to be measured is depen dent on the frequency as well as on the applied measure ment voltage The measuring instrument determines the values fo...

Страница 27: ...de termining the leakage inductance requires the secondary transformer side to be short circuited see fig 8 11 If the secondary side is short circuited the measured values of the primary side correspo...

Страница 28: ...easurement se tup are implemented it becomes necessary to perform a new compensation For more information about the OPEN SHORT compensation see chapter 7 2 8 5 Sorting Components with Option HO118 Bin...

Страница 29: ...nge for the sorting bins 0 5 and only the secondary parameters exceed the limit for sorting bin 6 7 General Fai lure BIN This sorting bin output is activated if the sorting does not apply to one of th...

Страница 30: ...sponds to TTL level and triggers with falling slopes It is protected against voltages of up to 15 volts For more information on the binning interface in the con text of the PIN and jumper assignment r...

Страница 31: ...E 488 controller control unit of a IEEE 488 bus system Any PC equipped with the corresponding plug in card can function as IEEE 488 controller If a IEC 625 cable must be used an appropriate plug in ad...

Страница 32: ...ort parallel processing of commands measuring instrument and if the instrument is ready to per form another measurement WAI The WAI command is a synchronization command that stops every subsequent com...

Страница 33: ...easurement circuit By de fault the automatic circuit type i 2 is selected Use i 0 The constant voltage CST V function must be switched on for measurements with bias current or external preload The err...

Страница 34: ...ative measurement value deviation REF_M for the main measurement value display if the measurement value de viation of the main measurement value display was previ ously activated via OUTP 1 or OUTP 2...

Страница 35: ...o 0 or if the automatic measurement mode AUTO is selected an error message is issued 10 3 Command List Binning Interface only with integrated binning interface HO118 XBIN The XBIN query queries the nu...

Страница 36: ...to 180 999 99 to 999 99 M 1 H to 100H N 0 95 to 500 Measurement conditions and functions Test frequency 20Hz to 200kHz 69 steps Frequency accuracy 100ppm AC test signal level 50mVrms to 1 5Vrms Resol...

Отзывы: