background image

Subject to change without notice

8

Basics of signal voltage

Type of signal voltage

The oscilloscope

 HM2005 

allows examination of DC voltages and

most repetitive signals in the frequency range up to at least
200MHz (-3dB). The vertical amplifiers have been designed for
minimum overshoot and therefore permit a true signal display.
The display of sinusoidal signals within the bandwidth limits
causes no problems, but an increasing error in measurement
due to gain reduction must be taken into account when measuring
high frequency signals. These errors become noticeable at approx.

100MHz

. At approx. 120MHz the reduction is approx. 10% and

the real voltage value is 11% higher. The gain reduction error can
not be defined exactly as the -3dB bandwidth of the amplifiers
differ between

 200MHz 

and

 220MHz

.

For sine wave signals the -6dB limits are approx. 280MHz.

When examining square or pulse type waveforms, attention must
be paid to the harmonic content of such signals. The repetition
frequency (fundamental frequency) of the signal must therefore
be significantly smaller than the upper limit frequency of the
vertical amplifier.
Displaying composite signals can be difficult, especially if they
contain no repetitive higher amplitude content which can be used
for triggering. This is the case with bursts, for instance. To obtain
a well triggered display in this case, the assistance of the variable
holdoff function or the second time base may be required.
Television video signals are relatively easy to trigger using the
built in TV-Sync-Separator (TV).

For optional operation as a DC or AC voltage amplifier, each vertical
amplifier input is provided with a DC/AC switch. DC coupling
should only be used with a series connected attenuator probe or
at very low frequencies or if the measurement of the DC voltage
content of the signal is absolutely necessary.

When displaying very low frequency pulses, the flat tops may be
sloping with AC coupling of the vertical amplifier (AC limit frequency
approx. 1.6 Hz for 3dB). In this case, DC operation is preferred,
provided the signal voltage is not superimposed on a too high DC
level. Otherwise a capacitor of adequate capacitance must be
connected to the input of the vertical amplifier with DC coupling.
This capacitor must have a sufficiently high breakdown voltage rating.
DC coupling is also recommended for the display of logic and pulse
signals, especially if the pulse duty factor changes constantly.
Otherwise the display will move upwards or downwards at each
change. Pure direct voltages can only be measured with DC-coupling.

The input coupling is selectable by the AC/DC pushbutton. The
actual setting is displayed in the readout with the

 “ = “

 symbol

for DC- and the

 “ ~ “ 

symbol for AC coupling.

Amplitude Measurements

In general electrical engineering, alternating voltage data normally
refers to effective values (rms = root-mean-square value).
However, for signal magnitudes and voltage designations in
oscilloscope measurements, the peak-to-peak voltage (V

pp

) value

is applied. The latter corresponds to the real potential difference
between the most positive and most negative points of a signal
waveform.
If a sinusoidal waveform, displayed on the oscilloscope screen,
is to be converted into an effective (rms) value, the resulting peak-

to-peak value must be divided by 2x

2 = 2.83. Conversely, it

should be observed that sinusoidal voltages indicated in Vrms
(Veff) have 2.83 times the potential difference in Vpp. The
relationship between the different voltage magnitudes can be
seen from the following figure.

Voltage values of a sine curve

V

rms

 = effective value; V

p

 = simple peak or crest value;

V

pp

 = peak-to-peak value; V

mom

 = momentary value.

The minimum signal voltage which must be applied to the Y input
for a trace of 1div height is 1mV

pp

 (± 5%) when this deflection

coefficient is displayed on the screen (readout) and the vernier is
switched off (VAR-LED dark). However, smaller signals than this
may also be displayed. The deflection coefficients are indicated
in mV/div or V/div (peak-to-peak value).

The magnitude of the applied voltage is ascertained by multiplying
the selected deflection coefficient by the vertical display height
in div. If an attenuator probe x10 is used, a further multiplication
by a factor of 10 is required to ascertain the correct voltage value.

For exact amplitude measurements, the variable control (VAR)
must be set to its calibrated detent CAL position.

With the variable control activated the deflection sensitivity can
be reduced up to a ratio of 2.5 to 1 (

please note “Controls and

Readout”

). Therefore any intermediate value is possible within

the 1-2-5 sequence of the attenuator(s).

With direct connection to the vertical input, signals up to
100Vpp may be displayed (attenuator set to 5V/div,
variable control to 2.5:1).

With the designations

H

= display height in div,

U

= signal voltage in V

pp

 at the vertical input,

D

= deflection coefficient in V/div at attenuator switch,

the required value can be calculated from the two given quantities:

However, these three values are not freely selectable. They have
to be within the following limits (trigger threshold, accuracy of
reading):

H

between 0.5 and 8div, if possible 3.2 to 8div,

U

between 1mV

pp

 and 40V

pp

,

D

between 1mV/div and 5V/div in 1-2-5 sequence.

Basics of signal voltage

Содержание HM 2005

Страница 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Страница 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Страница 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Страница 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Страница 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Страница 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Страница 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Страница 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Страница 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Страница 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Страница 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Страница 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Страница 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Страница 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Страница 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Страница 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Страница 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Страница 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Страница 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Страница 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Страница 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Страница 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Страница 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Страница 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Страница 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Страница 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Страница 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Страница 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Страница 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Страница 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Страница 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Страница 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Страница 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Страница 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Страница 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Страница 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Отзывы: