background image

21

Subject to change without notice

slowly. If the knob is pressed to the full extent the cursor
moves fast. The knob automatically reverts to the center
position and the cursor shift stops, if it is released.

(41) CAL.  

–  Pushbutton and concentric socket.

A square wave signal of 0.2Vpp ±1% is available from the
socket for probe adjustment purposes. The signal frequency
depends on the pushbutton setting. If the pushbutton is
released it is approx. 1kHz and can be switched over
(depressed) to approx. 1MHz. The pulse duty factor may
deviate from 1:1 and the frequency is not calibrated.

(42) CT  

–  Pushbutton and banana jack.

Pressing the pushbutton switches the instrument over from
oscilloscope to component test mode and vice versa.

In component test mode, all controls are deactivated with
the exception of the

 CT 

button,

 AUTOSET (2) 

and 

INTENS

(4). 

All LEDs except

 “A” 

or

 “RO” 

(associated with the INTENS

knob) are dark. The readout displays only 

“CT” 

in this

condition.

One test lead is connected to the CT socket. The second
test lead uses the ground socket

 (31)

.

 Please note

“Component Tester”

.

The maximum test voltage is approx. 20Vpp under open

circuit conditions, while the max. test current under short
circuit condition is approx. 20mApp.

Menu

The instrument software contains a menu and submenus. The
menu allows changes to the default settings regarding the
instrument behavior during operation, and a calibration option.

The menu can be called by pressing and holding the 

MENU

pushbutton. Then the readout displays the headline

 “MAIN

MENU” 

with the submenus

 ”TEST & CALIBRATE” 

and

”SETUP”

.

The following controls are relevant in menu operation:

1.

SAVE

 

(7)

 and 

RECALL (7)

 pushbutton selection function.

Briefly pressing selects the submenu or an item within the
submenu.

2.

SAVE (7)

 pushbutton 

SET

 function.

Pressing and holding the 

SAVE

 pushbutton calls the previously

selected submenu or item within the submenu. In those cases
where the item is marked with 

ON / OFF

 the setting changes

from 

ON

 to 

OFF

 or vice versa.

In some cases the called function is not performed at once
and a warning is displayed to protect from calling the function
inadvertently. Then the function can be called by pressing and
holding the 

SAVE

 pushbutton (

SET function

) again, otherwise

if the function was called inadvertently the proceeding can be
cancelled by pressing the 

AUTOSET (3)

 pushbutton.

3.

AUTOSET (3)

 pushbutton return function.

Each time the 

AUTOSET

 pushbutton is pressed, the menu is

switched back one step until 

MAIN MENU

 is displayed. Then

pressing the 

AUTOSET

 pushbutton again switches the menu

operation off and the 

AUTOSET

 pushbutton is reset to the

normal function.

The following menus, submenus and items within the submenus
are available:

1. MAIN MENU

1.1 TEST & CALIBRATE

Information regarding this can be found in the section ”Service
Instructions” item ”Adjustments”.

1.2 SETUP

This menu allows changes to the default settings regarding the
instrument behavior during operation. The SETUP menu contains
the submenus 

”MISCELLANEOUS”

 and 

”FACTORY”

.

1.2.1

MISCELLANEOUS

 contains:

1.2.1.1 CONTROL BEEP ON/OFF

In OFF condition the  acoustic signals actuated by the control
limits are switched off.

Note:  

The default setting is ON. If different conditions are required

the setting must be performed each time after switching the
oscilloscope on.

1.2.1.2 ERROR BEEP ON/OFF

Acoustic signals indicating faulty control operation are suppressed
in OFF condition.

Note: 

The default is ON. If different conditions are required the

setting must be performed each time after switching the
oscilloscope on.

1.2.1.3 QUICK START ON/OFF

In condition ON the HAMEG logo and the menus will not be
displayed after switching the instrument on. Then the instrument
is quickly ready for operation. To change from QUICK START OFF
to ON, press and hold the AUTOSET  button when switching the
instrument on until the menus become visible and change the
setting from ON to OFF.

1.2.1.4 TRIG.-  SYMBOL  ON/OFF

In most of the Yt (time base) modes the readout displays a trigger
point symbol which will not be displayed in condition OFF.

1.2.1.5 DC REFERENCE ON/OFF

If ON is selected and Yt (time base) mode is present, the readout
displays a ground symbol (). This symbol eases the evaluation
and determination of DC contents by indicating the 0 Volt
reference position.

1.2.1.6 INPUT Z ON/OFF

In condition ON the TRIG. EXT. / INP. (Z) BNC socket can be used
for trace intensity modulation. For further information please note
“Readout and Controls”.

1.2.1.7 MEAN VALUE ON/OFF

The mean value display is activated in the readout if ON is
selected and the cursor line measurement is switched off. For
further information please note section “Mean Value Display”.

1.2.2

FACTORY

 contains

Menu

Содержание HM 2005

Страница 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Страница 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Страница 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Страница 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Страница 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Страница 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Страница 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Страница 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Страница 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Страница 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Страница 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Страница 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Страница 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Страница 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Страница 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Страница 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Страница 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Страница 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Страница 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Страница 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Страница 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Страница 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Страница 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Страница 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Страница 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Страница 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Страница 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Страница 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Страница 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Страница 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Страница 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Страница 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Страница 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Страница 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Страница 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Страница 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Отзывы: