19
Subject to change without notice
Probe factor:
Pressing and holding the pushbutton selects the indicated
deflection coefficient of channel II displayed in the readout,
between 1:1 and 10:1. In condition 10:1 the probe factor is
thus indicated by a probe symbol displayed by the readout in
front the channel information (e.g. “probe symbol”, Y2...). In
the case of cursor voltage measurement, the probe factor is
automatically included.
Please note:
The symbol must not be activated unless a x10 (10:1)
attenuator probe is used.
(34) GD - INV.
– Pushbutton with two functions.
GD:
Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is displayed
in the readout as an earth (ground) symbol instead of the
deflection coefficient and the
~
(AC) or
=
(DC) (input coupling)
symbol.
The GD setting disables the input signal, the
AC/DC (33)
pushbutton and the
VOLTS/DIV (21)
knob. In automatic
trigger mode the undeflected trace is visible representing
the 0 Volt trace position.
The ”0 Volt” position is indicated by a ”
⊥
” symbol in the
readout. Please note
Y-POS. II (11)
.
INV.:
Pressing and holding this pushbutton switches the channel
II invert function on or off. The invert on condition is indicated
by the readout with a horizontal bar above
“Y2”
(Yt mode) or
“Y”
(XY mode). The invert function causes the signal display
of channel II to be inverted by 180°.
(35) TRIG. EXT. / INP. (Z)
– BNC socket with two functions.
The impedance of this input is 1M
Ω
II 20pF. The outer (ground)
connection is galvanically connected to the instrument
ground and consequently to the safety earth contact of the
line/mains plug.
TRIG. EXT.
:
Signals applied at this input can be used for external
triggering. Briefly pressing the
TRIG. (20)
pushbutton, until
the information
“TR:EXT, slope, coupling”
is visible in the
readout and the TRIG. “EXT” -LED is lit, switches the input
active as an external trigger source.
Z Input:
The input can be used for Z (trace intensity) modulation, if
software setting for “INPUT Z” is ON and neither external
triggering nor component tester operation is selected.
“INPUT Z” is part of the MISCELLANEOUS menu which is a
submenu of the SETUP menu.
The trace becomes blanked by a TTL high signal (positive
logic). Voltages higher then +5V are not permissible.
Below the CRT there are the controls for the readout, the
component tester and the square wave calibrator with their
outputs.
(36) MENU
– Pushbutton.
Pressing and holding this pushbutton activates the display
of the
MAIN MENU
. It contains the submenus
TEST &
CALIBRATE
and
SETUP
. The intensity of the display depends
on the
RO
-INTENS
(4)
setting.
Once a menu is displayed, the following pushbuttons are
of importance:
1. SAVE
and
RECALL (7)
pushbutton.
Briefly pressing selects the submenu or an item within
the submenu.
2. SAVE
–
SET (7)
pushbutton.
Pressing and holding the
SAVE (7)
pushbutton calls (sets)
the menu or the previously selected item. In those cases
where the item is marked with
ON / OFF
the setting
changes from
ON
to
OFF
or vice versa.
In some cases the called function is not performed at once
and a warning is displayed to protect from calling the
function inadvertently. Then the function can be called by
pressing and holding the
SAVE
pushbutton (
SET function
)
again, otherwise if the function was called inadvertently
the proceeding can be cancelled by pressing the
AUTOSET (3)
pushbutton.
3. AUTOSET (3)
pushbutton.
Each time the
AUTOSET
pushbutton is pressed the menu
is switched back one step until
MAIN MENU
is displayed.
Then pressing the
AUTOSET
pushbutton again switches
the menu operation off and the
AUTOSET
pushbutton
is
automatically set to the normal function.
(37) ON/OFF - CHI/II - 1/
∆∆∆∆∆
t
– This pushbutton has several
functions.
The following description of the cursor related controls
assumes that the readout is visible.
ON/OFF:
Pressing and holding the pushbutton switches both cursor
lines on or off. As the cursor lines are part of the readout,
they are visible only if the readout is switched on. If the cursor
lines are switched off and MEAN VALUE
ON
is activated
(Menu: SETUP > MISCELLANEOUS) the instrument is
prepared for DC Mean Value display. Please note section
“Mean Value Display”.
CHI/II:
This function
is required and available only in DUAL
and
XY
mode, if cursor supported voltage measurement (
∆∆∆∆∆
V (39)
)
is chosen, as the deflection coefficients may be different.
Briefly pressing the pushbutton switches between the
deflection coefficients of channel I and channel II and vice
versa. The measured result is displayed by the readout with
“
∆∆∆∆∆
V1:...”
or
“
∆∆∆∆∆
V2:...”
providing the deflection coefficient is
Controls and Readout
MADE IN GERMANY
INV.
INPUT CH I (HOR. INP.(X))
AC/DC
max.
250 Vp
1M
Ω
II
15pF
INPUT CH II
max.
250 Vp
1M
Ω
II
15pF
TRIG. EXT. INP
max.
100 Vp
(Z)
x1/x10
INV.
x1/x10
CAT
I
!
!
!
GD
AC/DC
GD
28
30
31
32
34
29
33
35
CT
ON
OFF
∆
V
∆
t
1
∆
t
1kHz
1MHz
CAL.
0.2 Vpp
C U R S O R
CH I/II
I/II
MENU
TRK
37
41
42
40
39 38 37
36