Probe factor:

Pressing and holding the pushbutton selects the indicated
deflection coefficient of channel II displayed in the readout,
between 1:1 and 10:1. In condition 10:1 the probe factor is
thus indicated by a probe symbol displayed by the readout in
front the channel information (e.g. “probe symbol”, Y2...).  In
the case of cursor voltage measurement, the probe factor is
automatically included.

Please note:
The symbol must not be activated unless a x10 (10:1)
attenuator probe is used.

(34) GD - INV. 

 –  Pushbutton with two functions.

GD:

Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is displayed
in the readout as an earth (ground) symbol instead of the
deflection coefficient and the

 ~ 

(AC) or

 =

 (DC) (input coupling)

symbol.

The GD setting disables the input signal, the

 AC/DC (33)

pushbutton and the

 VOLTS/DIV  (21) 

knob. In automatic

trigger mode the undeflected trace is visible representing
the 0 Volt trace position.

The ”0 Volt” position is indicated by a ”

” symbol in the

readout. Please note 

Y-POS. II (11)

.

INV.:

Pressing and holding this pushbutton switches the channel
II invert function on or off. The invert on condition is indicated
by the readout with a horizontal bar above 

“Y2” 

(Yt mode) or

“Y”

 (XY mode). The invert function causes the signal display

of channel II to be inverted by 180°.

(35) TRIG. EXT. / INP. (Z)

  –  BNC socket with two functions.

The impedance of this input is 1M

 II 20pF. The outer (ground)

connection is galvanically connected to the instrument
ground and consequently to the safety earth contact of the
line/mains plug.

TRIG. EXT.

:

Signals applied at this input can be used for external
triggering. Briefly pressing the

 TRIG.  (20) 

pushbutton, until

the information 

“TR:EXT, slope, coupling” 

is visible in the

readout and the TRIG. “EXT” -LED is lit, switches the input
active as an external trigger source.

Z Input:

The input can be used for Z (trace intensity) modulation, if
software setting for “INPUT Z” is ON and neither external
triggering nor component tester operation is selected.
“INPUT Z” is part of the MISCELLANEOUS menu which is a
submenu of the SETUP menu.

The trace becomes blanked by a TTL high signal (positive
logic). Voltages higher then +5V are not permissible.

Below the CRT there are the controls for the readout, the
component tester and the square wave calibrator with their
outputs.

(36) MENU  

–  Pushbutton.

Pressing and holding this pushbutton activates the display
of the 

MAIN MENU

. It contains the submenus 

TEST &

CALIBRATE 

and 

SETUP

. The intensity of the display depends

on the 

RO

-INTENS 

(4)

 setting.

Once a menu is displayed, the following pushbuttons are
of importance:

1. SAVE 

and

 RECALL (7) 

pushbutton.

Briefly pressing selects the submenu or an item within

the submenu.

2. SAVE 

 SET (7)  

pushbutton.

Pressing and holding the

 SAVE (7) 

pushbutton calls (sets)

the menu or the previously selected item. In those cases
where the item is marked with

 ON / OFF 

the setting

changes from

 ON 

to

 OFF 

or vice versa.

In some cases the called function is not performed at once

and a warning is displayed to protect from calling the
function inadvertently. Then the function can be called by
pressing and holding the

 SAVE 

pushbutton (

SET function

)

again, otherwise if the function was called inadvertently
the proceeding can be cancelled by pressing the

AUTOSET (3) 

pushbutton.

3. AUTOSET (3) 

pushbutton.

Each time the

 AUTOSET 

pushbutton is pressed the menu

is switched back one step until

 MAIN MENU 

is displayed.

Then pressing the

 AUTOSET 

pushbutton again switches

the menu operation off and the 

AUTOSET 

pushbutton

 

is

automatically set to the normal function.

(37) ON/OFF - CHI/II  - 1/

∆∆∆∆∆

t

  –  This pushbutton has several

functions.

The following description of the cursor related controls
assumes that the readout is visible.

ON/OFF:

Pressing and holding the pushbutton switches both cursor
lines on or off. As the cursor lines are part of the readout,
they are visible only if the readout is switched on. If the cursor
lines are switched off and MEAN VALUE 

ON

 is activated

(Menu: SETUP > MISCELLANEOUS) the instrument is
prepared for DC Mean Value display. Please note section
“Mean Value Display”.

CHI/II:

This function 

is required and available only in DUAL 

and

XY 

mode, if cursor supported voltage measurement (

∆∆∆∆∆

V (39)

)

is chosen, as the deflection coefficients may be different.

Briefly pressing the pushbutton switches between the
deflection coefficients of channel I and channel II and vice
versa. The measured result is displayed by the readout with

∆∆∆∆∆

V1:...” 

or

 “

∆∆∆∆∆

V2:...” 

providing the deflection coefficient is

Controls and Readout

MADE IN GERMANY

INV.

INPUT  CH I  (HOR. INP.(X))

AC/DC

max.
250 Vp

1M

 II 

15pF

INPUT  CH II 

max.
250 Vp

1M

 II 

15pF

TRIG. EXT. INP

max.
100 Vp

(Z)

x1/x10

INV.

x1/x10

CAT

 

I

!

!

!

GD

AC/DC

GD

28

30

31

32

34

29

33

35

CT

ON

OFF

V

t

1

t

1kHz
1MHz

CAL.

0.2 Vpp

C U R S O R

CH I/II

I/II

MENU

TRK

37

41

42

40

39 38 37

36

Содержание HM 2005

Страница 1: ...TRIG EXT INP max 100 Vp Z x1 x10 INV x1 x10 200 MHz ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV...

Страница 2: ...Subject to change without notice 2 M A N U A L H A N D B U C H M A N U E L...

Страница 3: ...X Y Operation 24 Phase comparison with Lissajous figures 24 Phase difference measurement in DUAL mode Yt 24 Measurement of amplitude modulation 25 Triggering and Time Base 25 Automatic Peak value Trig...

Страница 4: ...radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT g...

Страница 5: ...y modulation max 5V TTL Line Voltage Power consumption 100 240V AC 10 approx 43Watt 50 60Hz Min Max ambient temperature 10 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Size W...

Страница 6: ...ve neutral have been tested against insulation breakdown with 2200V DC The mains line plug shall only be inserted in a socket outlet provided with a protective earth contact The protective action must...

Страница 7: ...arly failures are detected by this method In the case of shipments by post rail or carrier the original packing must be used Transport damages and damage due to gross negligence are not covered by the...

Страница 8: ...can only be measured with DC coupling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amp...

Страница 9: ...for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must b...

Страница 10: ...the rise time of the probe e g 1 4ns If ttot is greater than 16ns then ttot can be taken as the rise time of the pulse and calculation is unnecessary Calculation of the example in the figure above re...

Страница 11: ...refore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max 1200 V DC peak AC A capacitor of corresponding capacitance and voltage rating may be connecte...

Страница 12: ...can be adjusted with a small screwdriver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile...

Страница 13: ...nction is practically ineffective in this settings 11 Y POS II Control knob The vertical trace position of channel II can be set with this control knob In ADD addition mode both Y POS I and Y POS II c...

Страница 14: ...id of the X POS control As the X expansion results in a higher time base speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over The X magni...

Страница 15: ...e related controls the Y POS I 9 knob and INV 30 button are deactivated For X position alteration the X POS 15 knob can be used 20 TRIG ALT Pushbutton with double function for trigger source selection...

Страница 16: ...DC trigger coupling Both settings are indicated by the NM 12 and the DC TRIG MODE LED 23 The previous trigger settings regarding the A time base remain unchanged In some trigger modes such as alterna...

Страница 17: ...ce This sector indicates the signal part which is displayed by the B time base The intensified segment can be shifted horizontally by the DEL POS 24 control knob continuously if the B time base is ope...

Страница 18: ...ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of the line mains plug 29 AC DC Pushbutton with two functions Input coupling Briefly...

Страница 19: ...and the square wave calibrator with their outputs 36 MENU Pushbutton Pressing and holding this pushbutton activates the display of the MAIN MENU It contains the submenus TEST CALIBRATE and SETUP The i...

Страница 20: ...t be taken into account In case of x10 10 1 probes the probe factor can be automatically included see item 29 and 33 If a x100 100 1 probe is used switch to the probe symbol 10 1 off if displayed Then...

Страница 21: ...on was called inadvertently the proceeding can be cancelled by pressing the AUTOSET 3 pushbutton 3 AUTOSET 3 pushbutton return function Each time the AUTOSET pushbutton is pressed the menu is switched...

Страница 22: ...s should be used Particular care is required when a single spot is displayed as a very high intensity setting may cause damage to the fluorescent screen of the CRT Switching the oscilloscope off and o...

Страница 23: ...the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator output Whilst it is not difficult for an ex...

Страница 24: ...measuring the distances a and b on the screen is quite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no i...

Страница 25: ...ed oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Y CH I 20mV div AC TIME DIV 0 2ms div Triggering Normal with LEVEL setting internal or externa...

Страница 26: ...displayed threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger indicator LED lights or flashes The internal trigger t...

Страница 27: ...visible TVL On the 10 s div setting and line TV triggering selected approx 1 lines are visible Those lines originate from the odd and even fields at random The sync separator circuit also operates wit...

Страница 28: ...ot coinciding with the slope selection setting The trigger coupling selection can also be used in external triggering mode Trigger indicator TR The following description applies to the TR LED Please n...

Страница 29: ...ng this function is part of the section Controls and Readout paragraph AUTOSET 2 As also mentioned in that section all controls are electronically selected with the exception of the POWER and calibrat...

Страница 30: ...s and transistors resistors capacitors and inductors Certain tests can also be made to integrated circuits All these components can be tested individually or in circuit provided that it is unpowered T...

Страница 31: ...des Possibly only a small portion of the knee is visible Zener diodes always show their forward knee and depending on the test voltage their zener breakdown forms a second knee in the opposite directi...

Страница 32: ...lvanically connected with the oscilloscope and subsequently with protective safety earth potential Measurement on a high level reference potential is not permitted and endangers operator oscilloscope...

Страница 33: ...switched off the RM LED 3 is dark Please note A minimum time must elapse between the commands RM 1 remote on and RM 0 remote off and vice versa The time can be calculated with the formula tmin 2x 1 ba...

Страница 34: ...ANALOG OSCILLOSCOPE HM2005 Instruments TRK CAT I GD AC DC GD CH I CH II DEL POS ADD CHP VAR VAR DUAL X Y VOLTS DIV VOLTS DIV TIME DIV 5V 1mV 5V 1mV 0 5s 20ns TRIG MODE TRIG ALT CHI AC DC HF NR LF TVL...

Страница 35: ...35 Subject to change without notice M A N U A L H A N D B U C H M A N U E L...

Страница 36: ...0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTracers T...

Отзывы: