31
Procedure
1.
Place the laser on the table at a spot that
provides the ability to measure all the
axes from the same setup.
2.
Level the Pitch-Axis level via by using
the coarse and fine adjustments to line
up the edges of the bubble. Repeat the
process with the Roll-Axis level vial.
Now the red laser plane is level to earth
3.
Place Target #1 at a location next to the
laser and mark the outline of the
magnetic base for good repeatability.
Point the laser at the target window and
adjust the target height until the laser
beam is centered.