5
6
Manufacturer’s identification label on controller– gives the manufacturer’s
name and address, and the model, serial number, and date of manufacture
of the equipment.
Table 2:
Safety Label Meanings
2.4. Safety Interlocks
The TLS-EQ-77H is equipped with interlocks to prevent operation of the device when any of the following
conditions are present:
•
An external interlock is open
•
No bulb is installed
2.5. External Interlocks
External interlock pins are provided for the customer’s use. Any suitable normally-open contact or solid-
state switch can operate the interlock circuit. The contact or switch should be rated for 80mA minimum at
5VDC.
The interlock circuit must be connected to enable the operation of the unit. Should the interlock connection
open during operation or standby, the source is immediately disabled, and all light output from the aperture
ceases.
2.6. Correct disposal of the unit
When the TLS-EQ-77H has finally been removed from service, observe all local environmental regulations
for proper disposal.
3. Preparation of System
3.1. Unpacking guide
Start by inspecting all parts of the system for completeness and any damage incurred in shipping. The TLS-
EQ77H shipping box should contain:
EQ-77 Power Supply Controller unit
TLS-EQ77H Lamp House unit
Black interconnecting cable from Lamp House to Power Supply Controller (21-pin mixed D-sub).
EQ-99-RC Remote Control Module with interlock connector
I/O cable with 15 pin D- connector at each end
Monochromator power supply
If any part is missing or appears damaged, contact Energetiq immediately. Do not attempt to substitute any
parts. There are no user-serviceable parts inside the TLS-EQ-77H Lamp House.