Alarm Display
If a problem occurs with the unit, the alarm indictor corresponding to each trouble flashes and at the same time, the
alarm signal is output from the external control terminal.
If occurrence of a problem is indicated, immediately check its cause and take corrective measures.
Alarm Display
●
Shutter error
If the shutter does not open (or close) when the shutter open
(or close) operation is carried out,
SHUTTER ERR.
indi-
cator flashes.
→
The shutter unit is defective. Contact our office.
●
Lamp error
If the lamp does not light even when the lamp lighting opera-
tion is carried out,
LAMP ERR.
Indicator flashes.
→
Check the lamp for proper installation. If it still
does not light, replace it with a new one.
If the outgoing light intensity fails to reach its set level while
the optical feedback function is activated, it is judged as
lamp deterioration and
LAMP ERR.
Indicator flashes.
→
Lower the output intensity setting or replace the
lamp with a new one. After replacing the lamp,
make sure to adjust the output intensity level.
To turn off
LAMP ERR.
indicator, press UP and
DOWN buttons at the same time.
●
Overheating
If the temperature inside the unit rises excessively high,
OVER HEAT
indicator flashes. If the overheated status
continues more than 2 minutes, the lamp is forcibly turned off.
→
Check for the following problems.
• Has the cooling fan stopped?
• Is the air intake port blocked?
• Is ambient temperature high due to insufficient ex-
haust from the airtight enclosure?
* When the temperature inside the unit lowers, the
alarm will be cancelled.
●
Lamp change
When the accumulated lamp operation time reach-
es 4000 hours (L8253:3000 hours) (lamp lifetime),
LAMP CHANGE
indicator flashes.
→
The unit can be used for some time after this indi-
cator starts flashing, replace the lamp with a new
one as soon as possible.
After replacing the lamp, be sure to reset the accu-
mulated lamp operation time.
When the accumulated lamp operation time reach-
es 5000 hours, the lamp is forced to turn off for the
sake of safety. (Unless the accumulated lamp oper-
ation time is reset, the lamp will not turn on again.)
Preparation prior to Operation
■
Connecting the Light Guide
q
Loosen the light guide fixing screw on the rear panel’s light guide connection aperture.
w
Insert a light guide all the way into the aperture, then retighten the screw.
●
Do not touch the end face of the light guide. Dirt at-
tached on it will reduce the amount of transmitted light.
●
Always handle light guides with care, as they are deli-
cate and may break if dropped or if bent to an exces-
sive angle. Never bend to a radius of less than 10 cm.
●
Do not use the light guide with the protection cap
attached to the end of the light guide. The cap would
melt as the end face of the fiber is deteriorated by
heat, and cause the output intensity (volume of light)
to reduce. If it has become dirty, wipe it with a gauze
(soft cloth) soaked with alcohol.
●
Always turn off the power supply switch to the main
unit before connecting the power supply.
■
Connecting the Power Supply
• Always make sure this unit has been grounded before using it.
• The power cord supplied with this unit is intended for use with this unit only. Do not use it
for any other purpose.
●
Always turn off the power supply switch to the main unit before connecting the power supply.
CAUTION
●
Do not touch the end
face of the light guide
with bare hand. If it has
become dirty, wipe it
with a gauze (soft cloth)
soaked with alcohol.
Light guide connection aperture
Never bend to a
radius of less than
10 cm.
Light guide
Caution
Caution
Содержание LIGHTINGCURE LC8
Страница 2: ...2...
Страница 6: ...6 1 1 1 2 JIS C 7550 3 JIS T 8141 30 kV 15 1 MPa 10 4 MPa 40...
Страница 8: ...8 25 80 10 cm 10 cm L9566 L9588 10 cm 10 cm 10 cm 10 cm...
Страница 11: ...11 q OFF w e 1 MPa 10 r t y u...
Страница 35: ...35 1 1 1 2 IEC 2006 IEC 62471 EMI 30kV 15 1MPa 10 4MPa 40...
Страница 37: ...37 L9566 L9588 10cm 10cm 10cm 10cm 25 80 10cm 10cm...
Страница 40: ...40 q OFF w e r t y 1MPa 10 u...
Страница 54: ...54...
Страница 55: ...55...