Valve Cover Lock Failure
Explanation:
An issue has been detected with the valve module cover latch.
Corrective Action:
1. Clamp the red, yellow, and blue lines.
2. Check the tubing manifold placement.
3. Check the tubing placement in the valve module.
4. Close and latch the valve module cover.
5. Unclamp the lines.
6. Close the device cover.
If the problem persists, service is required.
163
Valve Cover Not Closed
Explanation:
The valve module cover latch is not closed.
Corrective Action:
1. Clamp the red, yellow, and blue lines.
2. Check the tubing manifold placement.
3. Check the tubing placement in the valve module.
4. Close and latch the valve module cover.
5. Unclamp the lines.
6. Close the device cover.
If the problem persists, service is required.
164
System Fault
Explanation:
An issue has been detected with the device memory during the power-on self-test (POST).
Corrective Action:
Restart the device.
If the problem persists, service is required.