190
Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
230
Software Version Error: GUI
Explanation:
An issue has been detected with the user interface software version during power-up.
Corrective Action:
Please service the device before continuing use.
231
Software Update Complete
Explanation:
The software has been successfully updated. The device must be restarted before continuing
with normal operation.
Corrective Action:
Restart the device.
232
Blue Line Valve Position Error
Explanation:
An issue has been detected with the blue line valve position. This could be the result of spilled
fluid or debris causing the valve to stick.
Corrective Action:
1. Inspect the area around the blue line valve and clean if necessary.
2. Touch
Continue
to continue.
If the problem persists, service is required.
Table 16, Event Messages
ID #
Event Message Text