Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
302
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•CLOSE the white air line clamp.
•CLOSE the blue PRP line clamp.
•OPEN the yellow PPP line clamp.
2. Agitate the blood bag.
3. Touch
Continue
to fill the bowl.
303
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•CLOSE the white air line clamp.
•CLOSE the blue PRP line clamp.
•OPEN the yellow PPP line clamp.
2. Touch
Continue
to collect PPP.
304
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•CLOSE the white air line clamp.
•OPEN the blue PRP line clamp.
•CLOSE the yellow PPP line clamp.
2. Touch
Continue
to collect PRP.
Table 16, Event Messages
ID #
Event Message Text