Chapter 11, Troubleshooting
171
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
128
Fluid Detected Early
Explanation:
The bowl optics have detected fluid in bowl earlier than expected for the 70 mL bowl.
NOTE:
This could occur if saline was connected prior to the yellow line being installed in the
valve module and saline inadvertently entered the bowl.
Corrective Action:
If a 70 mL bowl is installed:
1. Check the bowl and chuck adaptor for proper installation in the centrifuge.
2. Touch
Keep Bowl Size
.
OR
If a Latham bowl is installed:
1. Touch
Change Bowl Size
.
2. Select the new bowl size or scan the processing set barcode.
129
Fluid Detected Early
Explanation:
The bowl optics have detected fluid in bowl earlier than expected for the 125 mL bowl.
NOTE:
This could occur if saline was connected prior to the yellow line being installed in the
valve module and saline inadvertently entered the bowl.
Corrective Action:
If a 125 mL bowl is installed:
1. Check the bowl for proper installation in the centrifuge.
2. Touch
Keep Bowl Size
.
OR
If a 70 mL bowl is installed:
1. Check the bowl and chuck adaptor for proper installation in the centrifuge.
2. Touch
Change Bowl Size
.
3. Select the new bowl size or scan the processing set barcode.
Table 16, Event Messages
ID #
Event Message Text