108
Air Detected in Yellow Line
Explanation:
The air detector sensed air in the yellow line. The saline bag may be empty.
Corrective Action:
1. Replace the saline bag if it is empty.
2. Ensure the yellow line clamp is open.
3. Check the tubing for kinks, occlusions, and proper placement in the air detector and pump.
4. Touch
Continue
to continue.
110
Air Detected During Conc
Explanation:
The air detector sensed air in the blue line during the Concentrate phase. The RBC bag may be
empty.
Corrective Action:
1. Ensure the blue line clamp is open.
2. Check the tubing for kinks, occlusions, and proper placement in the air detector and pump.
3. Wait for additional fluid in the reservoir.
OR
Touch
Wash
to wash a partial bowl.
111
Bowl is Ready to Wash
Explanation:
The optics have detected red cells. The bowl should be full and is ready to transition to the
Wash phase.
Corrective Action:
1. Touch
Wash
to enter the Wash phase OR
2. Touch
to continue filling the bowl OR
3. Touch
Return
to return cells to the reservoir.
Table 16, Event Messages
ID #
Event Message Text