Figure 4 Remove the transport locks
4.4 Connect to carrier gas
D A N G E R
Fire and explosion hazard. Oil and grease can burn with explosive violence when there is
oxygen. Make sure that all parts that are in contact with oxygen do not have grease or
oil.
The instrument cannot operate without a carrier gas connection. An oxygen or synthetic air supply
must be within 2 m (6.5 ft) of the rear of the analyzer. Use the supplied gas inlet tubing to connect
oxygen or synthetic air. Refer to
on page 3 for the gas supply specifications.
Connect the analyzer to a carrier gas as follows:
1.
Open the gas air supply for some seconds to remove unwanted material.
2.
Use a wrench to connect the supplied PTFE blue-stripe tubing to the gas supply and the Oxygen
port. Refer to
.
3.
Do not open the gas supply. Make sure that the carrier gas connection is leak tight.
English
11