Table 1 Reagents and standards
Chemical
Specification
Ultra-pure water (UPW)
Electrical conductivity at 25°C: 0.0555 μS/cm
maximum; electrical resistivity at 25°C:
18 MΩ·cm minimum
Note: UPW is used to prepare standards or dillutions. 2.5 L of
UPW is necessary for the analyzer installation.
Phosphoric acid (H
3
PO
4
) 85% w/w
3
Used to prepare the 9.33% phosphoric acid
solution (1 mol/L)
Note: The recommended acid solution is phosphoric acid
(H3PO4). As an alternative, other strong acids can be used.
Phosphoric acid solution (H
3
PO
4
) (1 mol/L)
9.33% w/w
54.9 mL is used to prepare 500 mL of solution.
Hydrochloric acid (HCl) 36% w/w
Used to prepare the 10.4% hydrochloric acid
solution.
Hydrochloric acid solution (HCl) (3 mol/L) 10,4%
w/w
144.7 mL is used to prepare 500 mL of solution.
Sulfuric acid (H
2
SO
4
) 96% w/w
Used to prepare the 13.5% sulfuric acid solution
Sulfuric acid solution (H
2
SO
4
) (1,5 mol/L) 13,5%
w/w
70.3 mL is used to prepare 500 mL of solution.
4.3 Remove the transport locks
W A R N I N G
Personal injury hazard. Instruments or components are heavy. Use assistance to install
or move.
N O T I C E
Use the bottom of the analyzer to lift the analyzer. Do not use the plastic covers or the sampler arm
to lift the analyzer or damage to the analyzer can occur.
Remove the magnetic lock and transportation strips from the analyzer. Refer to
and the
steps that follow:
1.
Remove the white stirrer lock.
2.
Remove the white strip.
3.
Remove the red warning strip..
4.
Discard the red warning strip
5.
Install the white strip again.
3
Concentration of the solution in terms of percentage composition. x% w/w solution equals X
grams of solute dissolved in 100 grams of solution.
10
English