Advanced Operations
6.7.3 Optical checks
For each optical check, the measured results must be evaluated
against user-specific requirements. The check options do not
define tolerances.
1.
Press
Optical Checks
in the System Checks menu.
The Optical Checks menu contains programs for checking the
wavelength accuracy, stray light and photometric accuracy.
An optional test filter set (Verification Kit) (
Section 9 on page 119
)
containing 6 precision glass filters, target values, tolerances and
instructions is available and is recommended as an aid for carrying
out comprehensive in-house instrument checks.
6.7.3.1 Wavelength check
The Wavelength Check test is used to check wavelength accuracy
at 807 nm.
1.
Press
Wavelength Check
in the Optical Checks menu.
2.
Insert the adapter (A) for 10-mm rectangular cells in the cell
compartment #2 and insert the sample cuvette/cell (Neodym or
BG20/2) in the adapter. Close the cell compartment.
Press
Start
.
3.
The result is displayed and is to be compared with the
nominal/standard data (given in the quality control certificate) of
the sample cuvette/cell.
4.
Press
Cancel
to return to Optical Checks.
Содержание Portable Spectrophotometer DR 2800
Страница 2: ...2...
Страница 6: ...6 Table of Contents...
Страница 8: ...8 Specifications...
Страница 22: ...22 Installation...
Страница 116: ...116 Maintenance...
Страница 118: ...118 Troubleshooting...
Страница 120: ...120 Replacement Parts...
Страница 122: ...122 Contact Information...
Страница 124: ...124 Limited Warranty...